A comparative viral load evaluation for human immunodeficiency virus, hepatitis B virus and hepatitis C virus of the Roche cobas 5800, Abbott Alinity m and Hologic Panther

Adriaan Meyer iD, Suné Saayman iD

Vox Sanguinis · 02.11.2026

DOI: 10.1111/vox.70407 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Vox Sanguinis, online erschienen am 02.11.2026. Verfasst von Adriaan Meyer und 1 weiteren Autor:innen. Beteiligte Einrichtung: Centre of Cardio-Metabolic Research in Africa, Division of Medical Physiology, Department of Biomedical Sciences, Stellenbosch University, Cape Town,…. Die Arbeit nennt 15 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background and Objectives Viral load (VL) quantification for human immunodeficiency virus (HIV), hepatitis B virus (HBV) and hepatitis C virus (HCV) is essential for clinical monitoring, treatment response assessment and post‐donation interpretation. With the introduction of automated molecular platforms, comparative evaluations are required to assess inter‐platform agreement, bias and precision across the quantitative VL range. Materials and Methods A comparative study was conducted using 280 plasma samples from blood donors. VL measurements were performed using the Roche cobas 5800 system, Abbott Alinity m system and the relevant Aptima quantitative assays on the Hologic Panther system. Agreement between platforms was assessed using Deming regression, Passing–Bablok regression, Lin's concordance correlation coefficient and Bland–Altman analyses. Precision was evaluated using repeatability and reproducibility testing using coefficients of variation (CVs). Differences in mean VL measurements were assessed with one‐way analysis of variance (ANOVA).

Results

The platforms showed comparable log 10 VL and acceptable precision across all targets. Deming regression identified constant bias in comparisons and proportional bias in one HCV comparison. Passing–Bablok regression and Lin's concordance correlation coefficient further showed that agreement differed. ANOVA showed no significant differences for HBV and HCV, while HIV showed a significant difference in VL values.

Conclusion

The platforms demonstrated acceptable precision for quantitative HIV, HBV and HCV VL testing. Roche cobas 5800 system showed low CVs in several repeatability comparisons, while within‐laboratory reproducibility showed variability for HBV and HIV. Agreement analyses identified constant bias, proportional bias and variability in platform comparisons.

Schlagwörter

Stichwörter der Autor:innen

method comparisonmolecular diagnostic platformsnucleic acid testingviral load platform comparisonviral load quantification

Autor:innen und Institutionen

  • Centre of Cardio-Metabolic Research in Africa, Division of Medical Physiology, Department of Biomedical Sciences, Stellenbosch University, Cape Town, South Africa.
  • Department of Transfusion Virology, Western Cape Blood Service, Cape Town, South Africa.

ORCID-Profile: Adriaan Meyer, Suné Saayman

Zitieren

APA

Meyer, A., & Saayman, S. (2026). A comparative viral load evaluation for human immunodeficiency virus, hepatitis B virus and hepatitis C virus of the Roche cobas 5800, Abbott Alinity m and Hologic Panther. Vox Sanguinis. https://doi.org/10.1111/vox.70407

Vancouver

Meyer A, Saayman S. A comparative viral load evaluation for human immunodeficiency virus, hepatitis B virus and hepatitis C virus of the Roche cobas 5800, Abbott Alinity m and Hologic Panther. Vox Sanguinis. 2026. doi:10.1111/vox.70407

BibTeX
@article{Meyer2026,
  author = {Meyer, Adriaan and Saayman, Suné},
  title = {A comparative viral load evaluation for human immunodeficiency virus, hepatitis B virus and hepatitis C virus of the Roche cobas 5800, Abbott Alinity m and Hologic Panther},
  journal = {Vox Sanguinis},
  year = {2026},
  doi = {10.1111/vox.70407}
}
Kostenlos starten →