EUVIMEDEUVIMED entdecken
EvaluationsstudieOpen Access · CC BY 4.0

Rapid anti‐HPA‐1a antibody quantification with a Luminex bead‐based assay: A method evaluation

Klara Asplund Högelin iD, Emöke Deschmann, Petter Höglund, Agneta Wikman

Vox Sanguinis · 2026

DOI: 10.1111/vox.70180 ↗

Kurz gefasst

Evaluationsstudie in der Zeitschrift Vox Sanguinis, 2026. Verfasst von Klara Asplund Högelin und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Clinical Immunology and Transfusion Medicine, Karolinska University Hospital, Stockholm, Sweden. Hauptthemen laut MEDLINE: Antigens, Human Platelet, Thrombocytopenia, Neonatal Alloimmune, Isoantibodies. Die Arbeit nennt 23 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background and Objectives High concentration of human platelet antigen‐1a (HPA‐1a) antibodies is reported to be associated with severe foetal and neonatal alloimmune thrombocytopaenia (FNAIT). The gold standard for quantification of anti‐HPA‐1a antibodies is the monoclonal antibody immobilization of platelet antigen (MAIPA) assay, which is a laborious method performed in only a few reference laboratories. The aim of this study was to evaluate the performance of the commercially available bead‐based Luminex assay PakLx (Immucor) for quantitative measurement of anti‐HPA‐1a antibodies. Materials and Methods We analysed anti‐HPA‐1a antibody levels in plasma samples from 42 HPA‐1a‐negative women who had given birth to a child with thrombocytopaenia. Quantification of antibodies was performed with two different techniques: MAIPA analysed by spectrophotometry with results expressed in international units (IU)/mL, and PakLx analysed in the Luminex assay with results expressed as the mean fluorescence intensity (MFI).

Results

In the comparison of the two methods' ability to stratify a result as either positive or negative, PakLx demonstrated 97.6% agreement with the MAIPA assay, with positive and negative predictive values of 96.7% and 100%, respectively. The correlation of the MFI values from PakLx with IU / mL in MAIPA assay was high, with a correlation coefficient ( R 2 ) of 0.92. MFI values were converted into semi‐quantitative results: high, intermediate and low levels of anti‐ HPA ‐1a.

Conclusion

PakLx shows high agreement with the MAIPA assay and allows fast laboratory turnaround time for the determination of anti‐HPA‐1a antibody levels. The result may be of predictive value in clinical assessments.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Antigens, Human PlateletThrombocytopenia, Neonatal AlloimmuneIsoantibodiesHumansFemaleInfant, NewbornPregnancyImmunoassayIntegrin beta3Antibodies, Monoclonal

Stichwörter der Autor:innen

HPA‐1a antibodiesHPA‐1a antibody quantificationfoetal and neonatal alloimmune thrombocytopaeniahuman platelet antigenlaboratory diagnostics

Substanzen

Antigens, Human PlateletIsoantibodiesITGB3 protein, humanIntegrin beta3Antibodies, Monoclonal

Autor:innen und Institutionen

  • Department of Clinical Immunology and Transfusion Medicine, Karolinska University Hospital, Stockholm, Sweden.
  • Department of Laboratory Medicine, Karolinska Institutet, Stockholm, Sweden.
  • Department of Women's and Children's Health, Karolinska Institutet, Stockholm, Sweden.
  • Department of Neonatal Medicine, Karolinska University Hospital, Stockholm, Sweden.
  • Department of Medicine Huddinge, Karolinska Institutet, Stockholm, Sweden.

ORCID-Profile: Klara Asplund Högelin

Zitieren

APA

Asplund Högelin, K., Deschmann, E., Höglund, P., & Wikman, A. (2026). Rapid anti‐HPA‐1a antibody quantification with a Luminex bead‐based assay: A method evaluation. Vox Sanguinis. https://doi.org/10.1111/vox.70180

Vancouver

Asplund Högelin K, Deschmann E, Höglund P, Wikman A. Rapid anti‐HPA‐1a antibody quantification with a Luminex bead‐based assay: A method evaluation. Vox Sanguinis. 2026. doi:10.1111/vox.70180

BibTeX
@article{AsplundHgelin2026,
  author = {Asplund Högelin, Klara and Deschmann, Emöke and Höglund, Petter and Wikman, Agneta},
  title = {Rapid anti‐HPA‐1a antibody quantification with a Luminex bead‐based assay: A method evaluation},
  journal = {Vox Sanguinis},
  year = {2026},
  doi = {10.1111/vox.70180}
}
Kostenlos starten →