Comparison of the TONOVET Plus®, TonoVet®, and Tono‐Pen Vet™ tonometers in normal cats and cats with glaucoma
Veterinary Ophthalmology · 2023
DOI: 10.1111/vop.13123 ↗Kurz gefasst
Fachartikel in der Zeitschrift Veterinary Ophthalmology, 2023. Verfasst von Julie A Kiland und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Ophthalmology and Visual Sciences, School of Medicine and Public Health, University of Wisconsin, Madison, Wisconsin, USA. Hauptthemen laut MEDLINE: Glaucoma, Cat Diseases. Die Arbeit wurde bisher 11-mal zitiert (Crossref) und nennt 40 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: NEI NIH HHS, National Eye Institute, Research to Prevent Blindness.
Abstract (Originalsprache)
Objectives
To determine the accuracy, precision, and clinical applicability of the ICare® TONOVET Plus (TVP) in cats. Animals and Procedures IOP readings obtained with the TVP were compared to values obtained concurrently with the original TONOVET (TV01) and Tono‐Pen Vet™ (TP) in 12 normal cats (24 eyes) and 8 glaucomatous LTBP2‐ mutant cats (13 eyes) in vivo. Reproducibility of TVP readings was also assessed for three observers in the above cats. The anterior chambers of five different normal cat eyes were cannulated ex vivo. IOP was measured with the TVP, TV01, and TP at manometric IOPs ranging from 5 to 70 mmHg. Data were analyzed by linear regression, ANOVA and Bland–Altman plots. ANOVA was used to assess reproducibility of TVP readings obtained by different observers and an ANCOVA model controlled for variation of individual cats. p < .05 was considered significant.
Results
TVP values strongly correlated with TV01 values ( y = 1.045 x + 1.443, R 2 = .9667). The TP significantly underestimated IOP relative to the TVP and TV01, particularly at high IOP. IOP values obtained by 1 observer were significantly higher (~1 mmHg average) compared to the other 2 observers via ANCOVA analysis ( p = .0006479 and p = .0203). Relative to manometry, the TVP and TV01 were significantly more accurate ( p < .0001) and precise ( p < .0070) than the TP in ex vivo eyes.
Conclusions
IOP readings obtained with the TVP and TV01 are broadly interchangeable between models and between observers, but subtle differences may be important in a research context. TP readings vastly underestimate high IOP in feline glaucoma.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Ophthalmology and Visual Sciences, School of Medicine and Public Health, University of Wisconsin, Madison, Wisconsin, USA.
- Department of Surgical Sciences, School of Veterinary Medicine, University of Wisconsin, Madison, Wisconsin, USA.
- McPherson Eye Research Institute, University of Wisconsin, Madison, Wisconsin, USA.
ORCID-Profile: Julie A Kiland, Hannah M Terhaar, Gillian J McLellan
Förderung
- NEI NIH HHS
- National Eye Institute
- Research to Prevent Blindness
Zitieren
APA
Kiland, J. A., Terhaar, H. M., Walleck, H. E., Chen, N., McDaniel, K., & McLellan, G. J. (2023). Comparison of the TONOVET Plus®, TonoVet®, and Tono‐Pen Vet™ tonometers in normal cats and cats with glaucoma. Veterinary Ophthalmology. https://doi.org/10.1111/vop.13123
Vancouver
Kiland JA, Terhaar HM, Walleck HE, Chen N, McDaniel K, McLellan GJ. Comparison of the TONOVET Plus®, TonoVet®, and Tono‐Pen Vet™ tonometers in normal cats and cats with glaucoma. Veterinary Ophthalmology. 2023. doi:10.1111/vop.13123
BibTeX
@article{Kiland2023,
author = {Kiland, Julie A and Terhaar, Hannah M and Walleck, Hannah E and Chen, Nickolas and McDaniel, Kyle and McLellan, Gillian J},
title = {Comparison of the TONOVET Plus®, TonoVet®, and Tono‐Pen Vet™ tonometers in normal cats and cats with glaucoma},
journal = {Veterinary Ophthalmology},
year = {2023},
doi = {10.1111/vop.13123}
}
EUVIMED