Evaluation of a whole blood point‐of‐care coagulation analyzer in dogs

Elizabeth M Conroy iD, Bridget M Lyons, Amie Koenig iD

Journal of Veterinary Emergency and Critical Care · 2024

DOI: 10.1111/vec.13416 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Veterinary Emergency and Critical Care, 2024. Verfasst von Elizabeth M Conroy und 2 weiteren Autor:innen. Beteiligte Einrichtung: Department of Emergency and Critical Care, University of Georgia, Athens, Georgia, USA. Hauptthemen laut MEDLINE: Point-of-Care Systems. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 28 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objective

To compare the accuracy of a point‐of‐care coagulation analyzer (POCCA) with a reference laboratory coagulation analyzer (LabCA) and to evaluate for confounding factors that could alter the performance of the POCCA.

Design

Prospective, observational study.

Setting

Two university veterinary teaching hospitals. Animals Forty‐three client‐owned dogs undergoing coagulation testing between April 2020 and June 2021.

Methods

Samples were obtained from dogs undergoing coagulation testing as part of a diagnostic workup. Prothrombin time (PT) and activated partial thromboplastin time (aPTT) were measured on the POCCA and on the LabCA. PCV, platelet count, total plasma protein, hyperbilirubinemia, hemolysis, lipemia, and autoagglutination were recorded.

Results

Moderate correlation was seen for PT and strong correlation was seen for aPTT between the POCCA and the LabCA (PT: 0.59, P < 0.0001; aPTT: 0.71, P < 0.0001). The POCCA results were consistent with normal or hypocoagulable samples for 30 of 38 PT and 33 of 37 aPTT results, as identified by the LabCA. Samples with PCV of 30%–55% were moderately correlated (PT: 0.63, P = 0.0004; aPTT: 0.63, P = 0.0003), but those outside that range were more likely to register an error message on the POCCA or provide disparate results. When hemolysis was present, there was a weak correlation between the POCCA and the LabCA for PT (rho: 0.38 [95% confidence interval: 0.19–0.76], P = 0.18) and a strong correlation for aPTT (rho: 0.86 [95% confidence interval: 0.62–0.95], P < 0.0001). Samples with hyperbilirubinemia were strongly correlated for PT (0.97, P = 0.002) but not for aPTT. Lipemia and autoagglutination were not observed.

Conclusion

There was an acceptable correlation in patients with PCV within the manufacturer's recommended reference range; however, measurements on samples with PCV outside the reference range were inconsistent with the LabCA. Caution should be used when using the POCCA in patients with coagulopathy and anemia or other potential confounders.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Point-of-Care SystemsAnimalsDogsProspective StudiesBlood Coagulation TestsFemaleMalePartial Thromboplastin TimeDog DiseasesProthrombin Time

Stichwörter der Autor:innen

bleedingcaninecoagulationhemostasispoint of caretrauma

Autor:innen und Institutionen

  • Department of Emergency and Critical Care, University of Georgia, Athens, Georgia, USA.
  • Department of Emergency and Critical Care, University of Missouri, Columbia, Missouri, USA.

ORCID-Profile: Elizabeth M Conroy, Amie Koenig

Zitieren

APA

Conroy, E. M., Lyons, B. M., & Koenig, A. (2024). Evaluation of a whole blood point‐of‐care coagulation analyzer in dogs. Journal of Veterinary Emergency and Critical Care. https://doi.org/10.1111/vec.13416

Vancouver

Conroy EM, Lyons BM, Koenig A. Evaluation of a whole blood point‐of‐care coagulation analyzer in dogs. Journal of Veterinary Emergency and Critical Care. 2024. doi:10.1111/vec.13416

BibTeX
@article{Conroy2024,
  author = {Conroy, Elizabeth M and Lyons, Bridget M and Koenig, Amie},
  title = {Evaluation of a whole blood point‐of‐care coagulation analyzer in dogs},
  journal = {Journal of Veterinary Emergency and Critical Care},
  year = {2024},
  doi = {10.1111/vec.13416}
}
Kostenlos starten →