EUVIMEDEUVIMED entdecken
Crossref

Validation of a New Quick and High‐Throughput Latex Agglutination Immuno‐Assay for Feline Alpha‐1‐Acid Glycoprotein

Kristina Museux iD, Paul Remmel, Jessy Leloup, Gregory Bouillet, Emma Montaufray, Maria Mateos, Aude Morgane Canonne

Veterinary Clinical Pathology · 08.11.2026

DOI: 10.1111/vcp.70162 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Veterinary Clinical Pathology, online erschienen am 08.11.2026. Verfasst von Kristina Museux und 6 weiteren Autor:innen. Beteiligte Einrichtung: Antech Diagnostics, Mars Petcare Science & Diagnostics, Massy, France. Die Arbeit nennt 36 Literaturangaben. Gefördert von: Antech Diagnostics, Mars Petcare Science and Diagnostics.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Alpha‐1‐acid glycoprotein (AGP) is currently considered the most sensitive and specific inflammatory marker for diagnosing feline infectious peritonitis (FIP) and monitoring cats undergoing antiviral treatment. Various feline AGP assays have been validated, but currently, no assay compatible with biochemistry analyzers is available for the rapid analysis of high sample volumes in reference laboratories.

Objectives

This study aimed to validate a novel method of AGP measurement in cats using a latex agglutination immunoassay on the Cobas 6000 (Roche, Meylan, France) and establish a reference interval (RI).

Methods

Surplus serum/heparin plasma sample pools with low (~330 μg/mL), medium (~970 μg/mL), and high (~2600 μg/mL) AGP concentrations, determined by the latex agglutination immunoassay, were used to establish intra‐ and inter‐assay coefficient of variation (CV). Precision, linearity, recovery, and stability were assessed using pooled samples. RI was determined on surplus serum samples of 43 healthy cats.

Results

The intra‐assay CV was 7.6%, 6.1%, and 3.5%, and the inter‐assay CV was 6.4%, 5.1%, and 6.2% for low, medium, and high AGP concentrations, respectively. Excellent linearity ( R 2 = 0.99) was observed for AGP concentrations between 194.1 and 1412.4 μg/mL. The upper reference limit was 622.8 μg/mL (90% confidence interval 517.1–713.7 μg/mL). Age and sex did not have a significant impact on AGP concentrations.

Conclusions

The latex agglutination immunoassay allows a precise and high‐throughput analysis of AGP concentrations in cats. This assay will greatly enhance the availability of AGP measurements.

Schlagwörter

Stichwörter der Autor:innen

AGPacute phase proteinscoronavirusfelinefeline infectious peritonitislatex agglutination immunoassay

Autor:innen und Institutionen

  • Antech Diagnostics, Mars Petcare Science & Diagnostics, Massy, France.
  • Internal Medicine Unit, National Veterinary School of Alfort, Maisons-Alfort, France.
  • Internal Medicine Unit, St Martin Veterinary Hospital Center, Allonzier-la-Caille, France.
  • Université Paris-Est Créteil, INSERM, IMRB, Créteil, France.

ORCID-Profile: Kristina Museux

Förderung

  • Antech Diagnostics, Mars Petcare Science and Diagnostics

Zitieren

APA

Museux, K., Remmel, P., Leloup, J., Bouillet, G., Montaufray, E., Mateos, M., & Canonne, A. M. (2026). Validation of a New Quick and High‐Throughput Latex Agglutination Immuno‐Assay for Feline Alpha‐1‐Acid Glycoprotein. Veterinary Clinical Pathology. https://doi.org/10.1111/vcp.70162

Vancouver

Museux K, Remmel P, Leloup J, Bouillet G, Montaufray E, Mateos M, et al. Validation of a New Quick and High‐Throughput Latex Agglutination Immuno‐Assay for Feline Alpha‐1‐Acid Glycoprotein. Veterinary Clinical Pathology. 2026. doi:10.1111/vcp.70162

BibTeX
@article{Museux2026,
  author = {Museux, Kristina and Remmel, Paul and Leloup, Jessy and Bouillet, Gregory and Montaufray, Emma and Mateos, Maria and Canonne, Aude Morgane},
  title = {Validation of a New Quick and High‐Throughput Latex Agglutination Immuno‐Assay for Feline Alpha‐1‐Acid Glycoprotein},
  journal = {Veterinary Clinical Pathology},
  year = {2026},
  doi = {10.1111/vcp.70162}
}
Kostenlos starten →