Agreement between hematocrit values determined by the Cobas b121 blood gas analyzer and the microhematocrit method in dogs, cats, and horses

Pâmela Dallarosa, Eduardo R Monteiro iD, Ana Paula S Borenstein, Stella F Valle iD

Veterinary Clinical Pathology · 2023

DOI: 10.1111/vcp.13245 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Veterinary Clinical Pathology, 2023. Verfasst von Pâmela Dallarosa und 3 weiteren Autor:innen. Beteiligte Einrichtung: Faculty of Veterinary Medicine, Federal University of Rio Grande do Sul, Porto Alegre, Brazil. Hauptthemen laut MEDLINE: Cat Diseases, Dog Diseases, Horse Diseases. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 11 Literaturangaben. Gefördert von: Conselho Nacional de Desenvolvimento Científico e Tecnológico, Universidade Federal do Rio Grande do Sul.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Packed cell volume (PCV) is important for assessing a patient's health status. Some blood gas analyzers measure hematocrit, and the agreement with PCV varies among different analyzers.

Objectives

We aimed to determine the agreement between PCV measured by microcentrifugation and hematocrit measured by the Cobas b121 blood gas analyzer in dogs, cats, and horses.

Methods

Whole blood samples for PCV and blood gas analysis were collected in lithium‐heparin syringes and analyzed within 10 min of collection. Agreement and association between the PCV and Cobas b121 generated hematocrit were assessed by the Bland–Altman method, Pearson's correlation, Deming regression analysis, and paired t tests. A total allowable error of 10% was used for the analysis.

Results

This study included 45 dogs, 45 cats, and 33 horses. The respective mean ± SD (minimum‐maximum) of PCVs and hematocrits were: dogs, 34.9 ± 9.9% (9.0–55.0) and 32.5 ± 8.8% (10.4–50.6); cats, 29.0 ± 9.6% (11.0–51.0) and 26.9 ± 9.3% (10.2–50.9); horses, 34.2 ± 6.5% (24.0–47.0) and 34.1 ± 6.0% (22.5–46.1). There were no significant differences between the methods. The bias ± SD was: dogs, −2.4 ± 2.6%; cats, −2.2 ± 2.3%; horses, −0.1 ± 2.4%. Pearson's correlation coefficients were > 0.90 for all species ( P < 0.0001). In 60%, 49%, and 85% of the samples for dogs, cats, and horses, respectively, the percentage differences between the methods were within 10%.

Conclusions

The Cobas b121 blood gas analyzer provided accurate estimates of PCVs in horses. However, in dogs and cats, there was a large frequency of unacceptable differences between the methods.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Cat DiseasesDog DiseasesHorse DiseasesCatsDogsHorsesAnimalsHematocritBlood Gas Analysis

Stichwörter der Autor:innen

blood conductivitycanineequinefelinepacked cell volume

Autor:innen und Institutionen

  • Faculty of Veterinary Medicine, Federal University of Rio Grande do Sul, Porto Alegre, Brazil.

ORCID-Profile: Eduardo R Monteiro, Stella F Valle

Förderung

  • Conselho Nacional de Desenvolvimento Científico e Tecnológico
  • Universidade Federal do Rio Grande do Sul

Zitieren

APA

Dallarosa, P., Monteiro, E. R., Borenstein, A. P. S., & Valle, S. F. (2023). Agreement between hematocrit values determined by the Cobas b121 blood gas analyzer and the microhematocrit method in dogs, cats, and horses. Veterinary Clinical Pathology. https://doi.org/10.1111/vcp.13245

Vancouver

Dallarosa P, Monteiro ER, Borenstein APS, Valle SF. Agreement between hematocrit values determined by the Cobas b121 blood gas analyzer and the microhematocrit method in dogs, cats, and horses. Veterinary Clinical Pathology. 2023. doi:10.1111/vcp.13245

BibTeX
@article{Dallarosa2023,
  author = {Dallarosa, Pâmela and Monteiro, Eduardo R and Borenstein, Ana Paula S and Valle, Stella F},
  title = {Agreement between hematocrit values determined by the Cobas b121 blood gas analyzer and the microhematocrit method in dogs, cats, and horses},
  journal = {Veterinary Clinical Pathology},
  year = {2023},
  doi = {10.1111/vcp.13245}
}
Kostenlos starten →