Validation of a blood gas device for ionized calcium analysis in Holstein cows
Veterinary Clinical Pathology · 2023
DOI: 10.1111/vcp.13233 ↗Kurz gefasst
Fachartikel in der Zeitschrift Veterinary Clinical Pathology, 2023. Verfasst von Abdülkerim Deniz und 3 weiteren Autor:innen. Beteiligte Einrichtung: Free Researcher for Clinical Biochemistry, İstanbul, Turkey. Hauptthemen laut MEDLINE: Hypocalcemia, Cattle Diseases. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 38 Literaturangaben.
Abstract (Originalsprache)
Background
Accurate analysis of ionized calcium (iCa) is critical for the detection of hypocalcemia or subclinical hypocalcemia. The Edan i15 Vet (EDAN) blood gas device has not been validated for iCa in dairy cows.
Objectives
We aimed to validate the EDAN blood gas device against the Gem Premier 3000 (GEM) analyzer by measuring iCa concentrations and evaluating the ability of these measurements, compared with serum total calcium (TCa) concentrations, to diagnose subclinical hypocalcemia. Methods iCa concentrations were measured with the EDAN and GEM devices, and serum TCa concentrations were measured with a wet biochemistry method with blood from 125 lactating Holstein cows between calving to day 27 postpartum.
Results
Bland‐Altman plots showed a mean and total bias of 0.05 and 0.24 mmol/L for the EDAN device, respectively. The intercept did not include zero, but the slope included 1.0 in the Passing‐Bablok regression. The sensitivity and specificity (Se/Sp) of the EDAN device were 93/94%, 93/90%, 91/93%, and 85/95% for iCa cut‐off values of
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Free Researcher for Clinical Biochemistry, İstanbul, Turkey.
- Department of Biochemistry, Faculty of Veterinary Medicine Milas, University of Muğla Sıtkı Koçman, Muğla, Turkey.
- Department of Internal Medicine, Faculty of Veterinary Medicine Milas, University of Muğla Sıtkı Koçman, Muğla, Turkey.
- Department of Statistics, Faculty of Science, University of Muğla Sıtkı Koçman, Muğla, Turkey.
ORCID-Profile: Abdülkerim Deniz, Mert Metin, Kemal Aksoy, Aytaç Pekmezci
Zitieren
APA
Deniz, A., Metin, M., Aksoy, K., & Pekmezci, A. (2023). Validation of a blood gas device for ionized calcium analysis in Holstein cows. Veterinary Clinical Pathology. https://doi.org/10.1111/vcp.13233
Vancouver
Deniz A, Metin M, Aksoy K, Pekmezci A. Validation of a blood gas device for ionized calcium analysis in Holstein cows. Veterinary Clinical Pathology. 2023. doi:10.1111/vcp.13233
BibTeX
@article{Deniz2023,
author = {Deniz, Abdülkerim and Metin, Mert and Aksoy, Kemal and Pekmezci, Aytaç},
title = {Validation of a blood gas device for ionized calcium analysis in Holstein cows},
journal = {Veterinary Clinical Pathology},
year = {2023},
doi = {10.1111/vcp.13233}
}
EUVIMED