Repeat patient testing‐quality control with canine samples shows promise as an alternative to commercial quality control material for a network of four Sysmex XT‐2000 i V hematology analyzers
Veterinary Clinical Pathology · 2024
DOI: 10.1111/vcp.13211 ↗Kurz gefasst
Fachartikel in der Zeitschrift Veterinary Clinical Pathology, 2024. Verfasst von Susan Daly und 2 weiteren Autor:innen. Beteiligte Einrichtung: SYNLAB-VPG/Cork, South Cork Industrial Estate, Cork, Ireland. Hauptthemen laut MEDLINE: Hemoglobins, Hematology. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 16 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background
Repeat patient testing‐quality control (RPT‐QC) uses retained patient samples as an alternative to commercial quality control material (QCM). We elected to calculate and validate RPT‐QC limits for red blood cell count (RBC), hemoglobin (HBG), hematocrit (HCT), and white blood cell count (WBC).
Objectives
(1) To validate RPT‐QC across a network of four harmonized Sysmex XT‐2000 i V hematology analyzers and determine the total error that can be controlled with RPT‐QC. (2) To generate quality control (QC) limits using the standard deviation (SD) of the duplicate measurement differences and determine a suitable simple QC rule with a probability of error detection >0.85 and probability of false rejection
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- SYNLAB-VPG/Cork, South Cork Industrial Estate, Cork, Ireland.
- School of Veterinary Medicine and Science, University of Nottingham, Leicestershire, UK.
- SYNLAB-VPG/Exeter, Exeter, UK.
ORCID-Profile: Susan Daly, Kathleen P Freeman
Zitieren
APA
Daly, S., Graham, P. A., & Freeman, K. P. (2024). Repeat patient testing‐quality control with canine samples shows promise as an alternative to commercial quality control material for a network of four Sysmex XT‐2000 i V hematology analyzers. Veterinary Clinical Pathology. https://doi.org/10.1111/vcp.13211
Vancouver
Daly S, Graham PA, Freeman KP. Repeat patient testing‐quality control with canine samples shows promise as an alternative to commercial quality control material for a network of four Sysmex XT‐2000 i V hematology analyzers. Veterinary Clinical Pathology. 2024. doi:10.1111/vcp.13211
BibTeX
@article{Daly2024,
author = {Daly, Susan and Graham, Peter A and Freeman, Kathleen P},
title = {Repeat patient testing‐quality control with canine samples shows promise as an alternative to commercial quality control material for a network of four Sysmex XT‐2000 i V hematology analyzers},
journal = {Veterinary Clinical Pathology},
year = {2024},
doi = {10.1111/vcp.13211}
}
EUVIMED