Long‐term stability and reproducibility of frozen plasma control samples for hepatitis B virus, hepatitis C virus, and human immunodeficiency virus by quantitative PCR : Implications for retrospective testing in hemovigilance
Transfusion · 2026-11 · 66 (11) · 2284-2292
DOI: 10.1111/trf.70398 ↗Kurz gefasst
Fachartikel in der Zeitschrift Transfusion, online erschienen am 19.09.2026. Verfasst von Benjamin Marx und 6 weiteren Autor:innen. Beteiligte Einrichtung: Institute for Medical Virology, University Hospital Frankfurt, Goethe University Frankfurt, Frankfurt am Main, Germany. Hauptthemen laut MEDLINE: Hepatitis B virus, Hepacivirus, HIV, Blood Safety. Die Arbeit nennt 16 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Objective
Data on the long‐term stability and reproducibility of nucleic acid amplification testing (NAT) results in frozen stored positive samples (≤−20°C) remain limited. This study evaluated the long‐term performance of frozen virus‐positive samples in quantitative PCR assays and aimed to provide evidence for their analytical reliability in retrospective NAT and quality assurance. Because blood transfusions, despite their high safety, still carry a residual risk of transmission of blood‐borne viruses, these data may also be relevant for look‐back investigations in hemovigilance. Materials and Methods Diluted and aliquoted internal run controls from positive patient samples were retrospectively analyzed using routine quantitative PCR assays. Long‐term stability was assessed over 37 months for HIV, 25 and 29 months for HBV‐A and HBV‐B, and 49 and 38 months for HCV‐A and HCV‐B. Descriptive statistics and linear regression were used to assess variability and temporal trends. Compliance with the German Medical Association's guidelines for quality assurance in medical laboratory testing (Rili‐BAEK) was evaluated using an acceptance range of ±0.5 log10.
Results
All internal run controls demonstrated stable long‐term performance during frozen storage and routine use. Across the datasets, 99.2%–100% of measurements remained within the acceptance range. Results outside mean ± 3 SD were rare, and no statistically significant temporal trend was observed for any internal run control.
Conclusion
Frozen HBV‐, HCV‐, and HIV‐positive samples showed robust long‐term stability and reproducibility in routine quantitative PCR assays for up to 4 years. These findings support the suitability of long‐term stored specimens for retrospective NAT.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Institute for Medical Virology, University Hospital Frankfurt, Goethe University Frankfurt, Frankfurt am Main, Germany.
- Fraunhofer Institute for Translational Medicine and Pharmacology (ITMP), Frankfurt am Main, Germany.
ORCID-Profile: Benjamin Marx, Annemarie Berger, Sebastian Hoehl, Holger F Rabenau
Zitieren
APA
Marx, B., Berger, A., Kohmer, N., Kortenbusch, M., Hoehl, S., Ciesek, S., & Rabenau, H. F. (2026). Long‐term stability and reproducibility of frozen plasma control samples for hepatitis B virus, hepatitis C virus, and human immunodeficiency virus by quantitative PCR : Implications for retrospective testing in hemovigilance. Transfusion, 66(11), 2284-2292. https://doi.org/10.1111/trf.70398
Vancouver
Marx B, Berger A, Kohmer N, Kortenbusch M, Hoehl S, Ciesek S, et al. Long‐term stability and reproducibility of frozen plasma control samples for hepatitis B virus, hepatitis C virus, and human immunodeficiency virus by quantitative PCR : Implications for retrospective testing in hemovigilance. Transfusion. 2026;66(11):2284-2292. doi:10.1111/trf.70398
BibTeX
@article{Marx2026,
author = {Marx, Benjamin and Berger, Annemarie and Kohmer, Niko and Kortenbusch, Marhild and Hoehl, Sebastian and Ciesek, Sandra and Rabenau, Holger F},
title = {Long‐term stability and reproducibility of frozen plasma control samples for hepatitis B virus, hepatitis C virus, and human immunodeficiency virus by quantitative PCR : Implications for retrospective testing in hemovigilance},
journal = {Transfusion},
year = {2026},
volume = {66},
number = {11},
pages = {2284--2292},
doi = {10.1111/trf.70398}
}
EUVIMED