Impact of input volume on red cell quality in deglycerolized RBCs using a modified ACP ‐215 protocol
Transfusion · 2025-01 · 65 (1) · 194-201
DOI: 10.1111/trf.18088 ↗Kurz gefasst
Studienprotokoll in der Zeitschrift Transfusion, online erschienen am 08.12.2024. Verfasst von Anita Howell und 8 weiteren Autor:innen. Beteiligte Einrichtung: Innovation and Portfolio Management, Canadian Blood Services, Edmonton, Alberta, Canada. Hauptthemen laut MEDLINE: Erythrocytes, Glycerol, Blood Preservation, Cryopreservation. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 14 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: Canadian Blood Services.
Abstract (Originalsprache)
Background
The ACP 215 automated cell processor is used to glycerolize and deglycerolize red cell concentrates (RCCs). Its primary advantage over the COBE 2991, previously used to cryopreserve RCCs, is that it maintains a closed system enabling extended post‐thaw expiry. However, it was observed that post‐deglycerolization hematocrits (Hct) of units processed with the LN236 kit are markedly lower than those processed using the COBE 2991. Therefore, we intended to determine whether a modified process using a smaller volume deglycerolization kit (LN235) could increase the final Hct with limited deleterious effects on product characteristics. Study Design and Methods Two proof‐of‐concept (POC) studies, conducted to determine the feasibility of using the LN235 processing kit for deglycerolization, identified the necessary modifications to the pre‐ and post‐deglycerolization process, after which a two‐part study characterized the modified protocol. The impact of pre‐cryopreservation storage duration (7–21 days), input red cell mass, and the type of CPD/SAGM RCC production method (red cell filtration and whole blood filtration) were investigated.
Results
Using the LN235 kit in conjunction with a volume reduction step for RCCs with a red cell mass exceeding 180 mL allowed for an ~8% increase in Hct. As expected, slightly lower recoveries were seen for large RCCs due to volume reduction; however, there were no other detrimental outcomes on product quality.
Conclusions
Leveraging the LN235 kit, recommended by Haemonetics for units with a red cell mass of ≤180 mL, can be used to increase the post‐deglycerolization Hct of RCCs that exceed this volume.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Innovation and Portfolio Management, Canadian Blood Services, Edmonton, Alberta, Canada.
- Supply Chain Process Management, Canadian Blood Services, Dartmouth, Nova Scotia, Canada.
- Supply Chain Process Management, Canadian Blood Services, Edmonton, Alberta, Canada.
- Innovation and Portfolio Management, Canadian Blood Services, Ottawa, Alberta, Canada.
- Department of Laboratory Medicine and Pathology, University of Alberta, Edmonton, Alberta, Canada.
ORCID-Profile: Jason P Acker
Förderung
- Canadian Blood Services
Zitieren
APA
Howell, A., Hill, A., Lefresne, W., Dennis, B., Turner, T. R., Yi, Q. L., Olafson, C., William, N., & Acker, J. P. (2025). Impact of input volume on red cell quality in deglycerolized RBCs using a modified ACP ‐215 protocol. Transfusion, 65(1), 194-201. https://doi.org/10.1111/trf.18088
Vancouver
Howell A, Hill A, Lefresne W, Dennis B, Turner TR, Yi QL, et al. Impact of input volume on red cell quality in deglycerolized RBCs using a modified ACP ‐215 protocol. Transfusion. 2025;65(1):194-201. doi:10.1111/trf.18088
BibTeX
@article{Howell2025,
author = {Howell, Anita and Hill, Angela and Lefresne, Wanda and Dennis, Brandie and Turner, Tracey R and Yi, Qi-Long and Olafson, Carly and William, Nishaka and Acker, Jason P},
title = {Impact of input volume on red cell quality in deglycerolized RBCs using a modified ACP ‐215 protocol},
journal = {Transfusion},
year = {2025},
volume = {65},
number = {1},
pages = {194--201},
doi = {10.1111/trf.18088}
}
EUVIMED