Evaluationsstudie in der Zeitschrift Transfusion Medicine, 2026. Verfasst von Yuelong Xu und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Clinical and Regulatory Affairs, InTec PRODUCTS, Inc., Xiamen, China. Hauptthemen laut MEDLINE: ABO Blood-Group System, Rh-Hr Blood-Group System, Blood Grouping and Crossmatching, Reagent Kits, Diagnostic. Die Arbeit nennt 24 Literaturangaben. Gefördert von: Medical and Health Technology Development Plan Project of Shandong Province (202302071560).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objectives

Rapid and accurate identification of blood groups is the foundation of emergency blood support. We performed a complete assessment of a new solid‐phase kit for ABO forward grouping and RhD grouping (ABD Kit, InTec Products, Xiamen, China) on the analytical performance, which was compared with those of traditional methods.

Methods

We analysed 1260 clinical samples using the ABD Kit, including weakly agglutinated samples of A, B and D antigen, and compared the test results with those via Gel card, test tube and slide methods. We also validated the results of typing blood samples containing weak antigens using first‐generation gene sequencing.

Results

The ABO forward group and RhD group of 1260 samples was determined using ABD Kit, revealing that the inter‐batch repeatability rate of the kit was 100%. The results of the kit were compared with those of the gel card method, revealing that the detection accuracy of the kit was also 100%, which was confirmed by comparing the detection of weak antigen samples with the results of first‐generation gene sequencing. The accuracy of the test tube agglutination method was 100%. In contrast, the accuracy of the slide agglutination method was low, especially in the detection of weak A blood antigens (agglutination strength 1+), weak B blood antigens (agglutination strength 1+) and weak D blood antigens (agglutination strength 1+ or 2+).

Conclusions

The ABD Kit (InTec Products, Xiamen, China) showed high sensitivity, reproducibility and specificity, indicative of excellent analytical performance. It is a reliable, practical and promising solution for rapid and accurate identification of blood types.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

ABO Blood-Group SystemRh-Hr Blood-Group SystemBlood Grouping and CrossmatchingReagent Kits, DiagnosticHumansFemaleMale

Stichwörter der Autor:innen

ABD kitABO forward groupingRhD groupinganalytical performancesolid‐phase membrane

Substanzen

ABO Blood-Group SystemRh-Hr Blood-Group SystemRho(D) antigenReagent Kits, Diagnostic

Autor:innen und Institutionen

  • Department of Clinical and Regulatory Affairs, InTec PRODUCTS, Inc., Xiamen, China.
  • Department of Blood Transfusion, the Affiliated Hospital of Qingdao University, Qingdao, China.

ORCID-Profile: Ying Li

Förderung

  • Medical and Health Technology Development Plan Project of Shandong Province (202302071560)

Zitieren

APA

Xu, Y., Wang, R., Ge, D., Huang, X., Yang, Y., Meng, Z., Wang, H., & Li, Y. (2026). Evaluation of a new solid‐phase ABO and RhD blood grouping kit. Transfusion Medicine. https://doi.org/10.1111/tme.70032

Vancouver

Xu Y, Wang R, Ge D, Huang X, Yang Y, Meng Z, et al. Evaluation of a new solid‐phase ABO and RhD blood grouping kit. Transfusion Medicine. 2026. doi:10.1111/tme.70032

BibTeX
@article{Xu2026,
  author = {Xu, Yuelong and Wang, Rong and Ge, Dongmei and Huang, Xingyu and Yang, Yuanming and Meng, Zifan and Wang, Haiyan and Li, Ying},
  title = {Evaluation of a new solid‐phase ABO and RhD blood grouping kit},
  journal = {Transfusion Medicine},
  year = {2026},
  doi = {10.1111/tme.70032}
}
Kostenlos starten →