Validation of the Sysmex XN analyser and Blood Bank mode for the quality and safety of donor blood and transfusion products
Transfusion Medicine · 2024
DOI: 10.1111/tme.13028 ↗Kurz gefasst
Fachartikel in der Zeitschrift Transfusion Medicine, 2024. Verfasst von Ondrej Valina und 4 weiteren Autor:innen. Beteiligte Einrichtung: Haematology Department, Sysmex Europe SE, Norderstedt, Germany. Hauptthemen laut MEDLINE: Blood Donors, Blood Banks. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 11 Literaturangaben.
Abstract (Originalsprache)
Objectives
Our objective was to compare the measurement of residual white blood cell (rWBC) and residual red blood cell (rRBC) counts in blood products using the XN Blood Bank mode and the laboratory standard operating procedures for manual counts. In addition, to compare the whole blood complete blood count (CBC) values of blood donors and the quality of blood products using the Sysmex XN analyser versus the XS‐1000i analyser. Materials and Methods For blood donors, 190 samples from blood or apheresis donors were analysed on both the Sysmex XS‐1000i and XN‐1000 analysers and the mean values of six CBC parameters were compared: the white blood cell count (WBC), the red blood cell count (RBC), haemoglobin (HGB), haematocrit (HCT), the mean corpuscular volume (MCV), the platelet count (PLT). For blood products, 164 samples were collected: 13 Plasma products – whole blood, 9 Plasma products – apheresis, 36 RBC concentrates – whole blood, 30 PLT concentrates – buffy coats, 36 PLT concentrates – buffy coats – pooled and 55 PLT concentrates – apheresis.
Results
All CBC parameters of the blood donors tested showed similar performance, with excellent correlation coefficients ( r ) ranging from 0.821 to 0.995. The majority of the blood products did not have a quantifiable number of residual cells, meaning the number of rWBC and rRBC, if present, was below the limit of quantitation (LoQ) of the different methods. rWBC were detected by Blood Bank mode in Plasma products – whole blood with a mean rWBC of 0.012 × 10 9 /L and in PLT concentrates – buffy coats with a mean rWBC of 0.19 × 10 9 /L. The correlation coefficient in both analysers for all three parameters (HGB, HCT, RBC) in RBC concentrates – whole blood was excellent, ranging from 0.95 to 0.99. For platelet count, r ranged from 0.98 to 0.99.
Conclusion
The XN‐Series analyser, equipped with a Blood Bank mode, demonstrated reliable performance when used for blood donor evaluation, rWBC enumeration and measurement of end blood products.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Haematology Department, Sysmex Europe SE, Norderstedt, Germany.
- Apheresis and Transfusion Departments, Institute of Haematology and Blood Transfusion, Prague, Czech Republic.
Zitieren
APA
Valina, O., Vankova, I., Mintzas, K., Knappova, K., & Gasova, Z. (2024). Validation of the Sysmex XN analyser and Blood Bank mode for the quality and safety of donor blood and transfusion products. Transfusion Medicine. https://doi.org/10.1111/tme.13028
Vancouver
Valina O, Vankova I, Mintzas K, Knappova K, Gasova Z. Validation of the Sysmex XN analyser and Blood Bank mode for the quality and safety of donor blood and transfusion products. Transfusion Medicine. 2024. doi:10.1111/tme.13028
BibTeX
@article{Valina2024,
author = {Valina, Ondrej and Vankova, Ida and Mintzas, Konstantinos and Knappova, Klara and Gasova, Zdenka},
title = {Validation of the Sysmex XN analyser and Blood Bank mode for the quality and safety of donor blood and transfusion products},
journal = {Transfusion Medicine},
year = {2024},
doi = {10.1111/tme.13028}
}
EUVIMED