Characterization and parameters for the effectiveness of a UV ‐C device for infection control in dental clinics
Photochemistry and Photobiology · 2026
DOI: 10.1111/php.70065 ↗Kurz gefasst
Fachartikel in der Zeitschrift Photochemistry and Photobiology, 2026. Verfasst von Kátia Cristiane Hall und 8 weiteren Autor:innen. Beteiligte Einrichtung: Graduate Program in Dentistry (PPGO), School of Dentistry Federal University of Pelotas (UFPel), Pelotas, RS, Brazil. Hauptthemen laut MEDLINE: Ultraviolet Rays, Dental Clinics, Disinfection. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 26 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Fundação de Amparo à Pesquisa do Estado do Rio Grande do Sul, Secretaria de Desenvolvimento Econômico, Ciência, Tecnologia e Inovação.
Themen: Nosokomiale Infektionen
Abstract (Originalsprache)
Healthcare‐associated infections (HAIs) remain a significant concern in dental settings, especially in light of increasing antimicrobial resistance. This study aimed to characterize and evaluate the efficacy of an ultraviolet C (UV‐C) disinfection device specifically developed as an additional resource for dental clinics and healthcare settings. Colorimetric dosimeters were strategically distributed at different points and vertical levels to monitor UV‐C radiation distribution in the environment. The emission spectrum and radiation intensity were measured using spectral measurement techniques, while microbiological analyses were conducted on seven commonly encountered surfaces in dental offices using mannitol salt agar selective for Staphylococcus aureus and Staphylococcus epidermidis . The results revealed a significant reduction in colony‐forming units (CFUs), exceeding 90% at most sampling points ( p < 0.05). An exponential falloff of irradiance with distance from the UV‐C source was observed, highlighting the importance of proper positioning of the equipment. Notably, the UV‐C device proved effective even on surfaces with higher microbial loads, such as the armrest of the dental chair. The findings demonstrate that both time and distance significantly affect disinfection efficacy, and that purpose‐built UV‐C devices are viable as complementary tools to conventional chemical cleaning and disinfection protocols.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Graduate Program in Dentistry (PPGO), School of Dentistry Federal University of Pelotas (UFPel), Pelotas, RS, Brazil.
- Graduate Program in Electronics and Computer Engineering (MEEC), Catholic University of Pelotas (UCPel), Pelotas, RS, Brazil.
- Graduate Program in Materials Science and Engineering (PPGCEM), Federal University of Pelotas (UFPel), Pelotas, RS, Brazil.
- Institute of Physics and Mathematics (IFM), Graduate Program in Physics (PPGFis), Federal University of Pelotas (UFPel), Capão do Leão, RS, Brazil.
ORCID-Profile: Everton Granemann Souza, Mário Lúcio Moreira, Rafael Guerra Lund, Evandro Piva
Förderung
- Fundação de Amparo à Pesquisa do Estado do Rio Grande do Sul
- Secretaria de Desenvolvimento Econômico, Ciência, Tecnologia e Inovação
Zitieren
APA
Hall, K. C., Souza, E. G., do Nascimento, C. D. D., Basso, M. L., Moreira, M. L., Pena, C. L. D., da Rosa, W. L. O., Lund, R. G., & Piva, E. (2026). Characterization and parameters for the effectiveness of a UV ‐C device for infection control in dental clinics. Photochemistry and Photobiology. https://doi.org/10.1111/php.70065
Vancouver
Hall KC, Souza EG, do Nascimento CDD, Basso ML, Moreira ML, Pena CLD, et al. Characterization and parameters for the effectiveness of a UV ‐C device for infection control in dental clinics. Photochemistry and Photobiology. 2026. doi:10.1111/php.70065
BibTeX
@article{Hall2026,
author = {Hall, Kátia Cristiane and Souza, Everton Granemann and do Nascimento, Chiara das Dores and Basso, Marcel Luiz and Moreira, Mário Lúcio and Pena, Carla Lucía David and da Rosa, Wellington Luiz Oliveira and Lund, Rafael Guerra and Piva, Evandro},
title = {Characterization and parameters for the effectiveness of a UV ‐C device for infection control in dental clinics},
journal = {Photochemistry and Photobiology},
year = {2026},
doi = {10.1111/php.70065}
}
EUVIMED