EUVIMEDEUVIMED entdecken
BeobachtungsstudieOpen Access · CC BY 4.0

Utility of non‐contact thermometers in neonatal intensive care: Effects of incubator conditions and measurement sites

Nützlichkeit berührungsloser Thermometer in der neonatologischen Intensivmedizin: Einfluss von Inkubatorbedingungen und Messorten

Takashi Okuno iD, Tatsuto Shimizu, Aiko Igarashi iD, Yusei Ohshima iD

Pediatrics International · 2025-01 · 67 (1)

DOI: 10.1111/ped.70231 ↗

Auf einen Blick

Bei Frühgeborenen korrelierten NCIT-Messungen mit der axillären CT-Temperatur, variierten aber je nach Messort. Am Rumpf waren die Übereinstimmung und im HT-Bereich die Limits of agreement am besten.

Fragestellung

Wie zuverlässig sind berührungslose Infrarotthermometer unter verschiedenen Temperaturbedingungen im Inkubator, und welche Messorte sind am geeignetsten?

Methode

Beobachtungsstudie bei Frühgeborenen mit korrigiertem Gestationsalter von 25–32 Wochen; Messung mit NCIT an Axilla, Stirn und Rumpf sowie CT in der Axilla; Vergleich in HT- und LT-Umgebungen.

Ergebnisse

NCIT-Temperaturen korrelierten mit der axillären CT-Temperatur, unterschieden sich aber nach Messort. Der berührungslose Rumpfwert zeigte im HT-Bereich die beste Korrelation und die kleinsten Limits of agreement; Mittelwerte mehrerer Orte verbesserten die Korrelation (p < 0,05).

Einschränkungen

Im LT-Bereich zeigten einzelne NCIT-Messorte breite Limits of agreement.

Schlussfolgerung der Autor:innen

NCITs scheinen für die Temperaturüberwachung bei Neugeborenen praktikabel zu sein, besonders an Rumpfoberflächen. Mittelwerte aus mehreren Messorten können die Genauigkeit verbessern.

Kernaussagen

  • NCITs korrelierten mit der axillären CT-Temperatur
  • Am Rumpf waren Übereinstimmung und Genauigkeit am besten
  • Mittelwerte aus mehreren Messorten verbesserten die Korrelation

Deutsche Zusammenfassung: KI-gestützt aus dem Original-Abstract erstellt; alle Zahlen wurden automatisch mit dem Original abgeglichen. Maßgeblich ist der Original-Abstract.

Kurz gefasst

Beobachtungsstudie in der Zeitschrift Pediatrics International, online erschienen am 14.10.2025. Verfasst von Takashi Okuno und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Pediatrics, Faculty of Medical Sciences, University of Fukui, Fukui, Japan. Hauptthemen laut MEDLINE: Incubators, Infant, Thermometers, Body Temperature, Intensive Care, Neonatal. Die Arbeit nennt 18 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

Auf einen BlickAbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Non‐contact infrared thermometers (NCITs) are non‐invasive alternatives to conventional contact thermometers (CTs). However, the accuracy of NCITs may be affected by measurement environments and body surface sites. This study evaluated the reliability of NCIT measurements under various temperature conditions in neonatal incubators and explored optimal measurement strategies.

Methods

Preterm neonates with a corrected gestational age of 25–32 weeks were enrolled in this observational study. Body temperature was measured using NCITs at the axilla, forehead, and both contact and non‐contact surfaces of the trunk, and by CT in the axilla. Measurement environments were classified into two groups: high temperature (HT [≥33°C, ≥60% humidity]) and low temperature (LT [28.5–32.9°C, ≥40% humidity]).

Results

Body temperatures measured by NCIT correlated with axillary temperature measured by CT (axillary CT) but varied by site. The non‐contact trunk surface measured by NCIT had the best correlation with temperature measured by axillary CT and had the smallest limits of agreement in the HT group. However, Bland–Altman analysis showed wide limits of agreement for single‐site NCITs in the LT group, although the correlation at the contact trunk surface improved with corrected gestational age. Averaging NCIT readings from the axilla and both trunk surfaces yielded stronger correlations with axillary CT ( p < 0.05).

Conclusions

NCITs appear to be a practical option for neonatal temperature monitoring, particularly when used at trunk surface sites. Taking average measurements from multiple sites may enhance accuracy and aid in creating better measurement protocols in NICUs.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Incubators, InfantThermometersBody TemperatureIntensive Care, NeonatalHumansInfant, NewbornFemaleMaleAxillaReproducibility of ResultsInfant, PrematureForeheadGestational AgeIntensive Care Units, Neonatal

Stichwörter der Autor:innen

body temperature regulationincubatorinfrared thermometernewborntemperature measurement

Autor:innen und Institutionen

  • Department of Pediatrics, Faculty of Medical Sciences, University of Fukui, Fukui, Japan.

ORCID-Profile: Takashi Okuno, Aiko Igarashi, Yusei Ohshima

Zitieren

APA

Okuno, T., Shimizu, T., Igarashi, A., & Ohshima, Y. (2025). Utility of non‐contact thermometers in neonatal intensive care: Effects of incubator conditions and measurement sites. Pediatrics International, 67(1). https://doi.org/10.1111/ped.70231

Vancouver

Okuno T, Shimizu T, Igarashi A, Ohshima Y. Utility of non‐contact thermometers in neonatal intensive care: Effects of incubator conditions and measurement sites. Pediatrics International. 2025;67(1). doi:10.1111/ped.70231

BibTeX
@article{Okuno2025,
  author = {Okuno, Takashi and Shimizu, Tatsuto and Igarashi, Aiko and Ohshima, Yusei},
  title = {Utility of non‐contact thermometers in neonatal intensive care: Effects of incubator conditions and measurement sites},
  journal = {Pediatrics International},
  year = {2025},
  volume = {67},
  number = {1},
  doi = {10.1111/ped.70231}
}
Kostenlos starten →