Comparison of tidal volume and airway pressure during neonatal resuscitation in mask leaks

Vergleich von Tidalvolumen und Atemwegsdruck bei neonataler Reanimation unter Maskenlecks

Fumihiko Takatori iD, Tetsuya Isayama, Shigeharu Hosono iD, Kiyotaka Iwasaki

Pediatrics International · 2025-01 · 67 (1)

DOI: 10.1111/ped.70139 ↗

Auf einen Blick

In einem Modell der neonatalen Reanimation unterschätzten RFM-Messungen das tatsächliche Tidalvolumen bei Maskenlecks. Der Unterschied nahm mit zunehmendem Leck zu.

Fragestellung

Unterscheiden sich Tidalvolumen und Atemwegsdruck zwischen Respiratory Function Monitor und künstlicher Lunge bei Maskenlecks?

Methode

Drei Experimente in einem System aus PPV-Gerät, Gesichtsmaske und künstlicher Lunge; Vergleich von RFM in Testlunge und Maske unter verschiedenen Lungenbedingungen, Ventilationsraten und PPV-Geräten.

Ergebnisse

Maskenlecks führten zur Unterschätzung des tatsächlichen Tidalvolumens; der Effekt nahm mit steigender Leckrate zu. Mit Druckgasquelle sank das Tidalvolumen von 15 auf 12 mL, ohne von 16 auf 5 mL.

Schlussfolgerung der Autor:innen

Es zeigten sich deutliche Abweichungen zwischen RFM-Werten und Testlunge beim Tidalvolumen. Eine genaue Überwachung ist wichtig, um Lungenschäden durch zu großes Tidalvolumen zu vermeiden.

Kernaussagen

  • Maskenlecks führten zur Unterschätzung des Tidalvolumens
  • Der Effekt nahm mit steigender Leckrate zu
  • Messabweichungen waren mit Druckgasquelle geringer

Deutsche Zusammenfassung: KI-gestützt aus dem Original-Abstract erstellt; alle Zahlen wurden automatisch mit dem Original abgeglichen. Maßgeblich ist der Original-Abstract.

Kurz gefasst

Vergleichsstudie in der Zeitschrift Pediatrics International, online erschienen am 02.07.2025. Verfasst von Fumihiko Takatori und 3 weiteren Autor:innen. Beteiligte Einrichtung: Cooperative Major in Advanced Biomedical Sciences, Joint Graduate School of Tokyo Women's Medical University and Waseda University, Waseda University…. Hauptthemen laut MEDLINE: Tidal Volume, Positive-Pressure Respiration, Masks, Resuscitation. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 14 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

Auf einen BlickAbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Providing positive‐pressure ventilation (PPV) to the lungs is essential for neonatal resuscitation. Accurate PPV requires a precise measurement of tidal volume and airway pressure, with airway obstruction and mask leakage being the primary concerns for ineffective ventilation. This study aimed to investigate the differences between tidal volume and airway pressure measured by a respiratory function monitor (RFM) and the actual values delivered to the lungs in scenarios involving mask leaks, using a system comprising a PPV device, a face mask, and an artificial lung model.

Methods

Three experiments were conducted to assess mask leakage (1) under varying lung conditions, (2) under different ventilation rates, and (3) using different PPV devices. Two RFMs were used, one in the test lung and the other in the mask. Trends in those data were assessed by means of a correlation graph.

Results

Mask leakage resulted in an underestimation of the actual tidal volume, with the effect intensifying as the leak percentage increased. PPV devices using a compressed gas source demonstrated less reduction in lung tidal volume (from 15 to 12 mL) owing to mask leaks compared with those without such a source (from 16 to 5 mL).

Conclusions

Significant discrepancies were observed between RFM readings and test lung values for tidal volume. These findings highlight the importance of accurate monitoring to prevent lung injury caused by excessive tidal volume, particularly in the presence of mask leaks. Accurately measuring tidal volume in the presence of mask leaks presents a significant challenge for the future.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Tidal VolumePositive-Pressure RespirationMasksResuscitationHumansInfant, NewbornEquipment FailureRespiratory Function Tests

Stichwörter der Autor:innen

infant newbornpositive‐pressure respirationresuscitation

Autor:innen und Institutionen

  • Cooperative Major in Advanced Biomedical Sciences, Joint Graduate School of Tokyo Women's Medical University and Waseda University, Waseda University, Tokyo, Japan.
  • Ventilator & Anesthesia Products Department, Nihon Kohden Corporation, Tokyo, Japan.
  • Division of Neonatology, National Center for Child Health and Development, Tokyo, Japan.
  • Division of Neonatology, Jichi Medical University Saitama Medical Center, Saitama, Japan.
  • Department of Modern Mechanical Engineering, School of Creative Science and Engineering, Waseda University, Tokyo, Japan.
  • Department of Integrative Bioscience and Biomedical Engineering, Graduate School of Advanced Science and Engineering, Waseda University, Tokyo, Japan.

ORCID-Profile: Fumihiko Takatori, Shigeharu Hosono

Zitieren

APA

Takatori, F., Isayama, T., Hosono, S., & Iwasaki, K. (2025). Comparison of tidal volume and airway pressure during neonatal resuscitation in mask leaks. Pediatrics International, 67(1). https://doi.org/10.1111/ped.70139

Vancouver

Takatori F, Isayama T, Hosono S, Iwasaki K. Comparison of tidal volume and airway pressure during neonatal resuscitation in mask leaks. Pediatrics International. 2025;67(1). doi:10.1111/ped.70139

BibTeX
@article{Takatori2025,
  author = {Takatori, Fumihiko and Isayama, Tetsuya and Hosono, Shigeharu and Iwasaki, Kiyotaka},
  title = {Comparison of tidal volume and airway pressure during neonatal resuscitation in mask leaks},
  journal = {Pediatrics International},
  year = {2025},
  volume = {67},
  number = {1},
  doi = {10.1111/ped.70139}
}
Kostenlos starten →