Use of real‐time respiratory function monitor improves neonatal face mask ventilation: Cross‐over simulation study

Prakash Kannan Loganathan iD, Charlotte Ashton, Emily Harrold, Sarah Wigston, Le Minh Thao Doan, Annalisa Occhipinti

Pediatric Anesthesia · 2025-01 · 35 (1) · 66-74

DOI: 10.1111/pan.15020 ↗

Kurz gefasst

Beobachtungsstudie in der Zeitschrift Pediatric Anesthesia, online erschienen am 07.10.2024. Verfasst von Prakash Kannan Loganathan und 5 weiteren Autor:innen. Beteiligte Einrichtung: Neonatal Unit, James Cook University Hospital, Middlesbrough, UK. Hauptthemen laut MEDLINE: Cross-Over Studies, Manikins, Masks, Tidal Volume. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 21 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Neonatal resuscitation airway training can be difficult as there is no feedback on the face mask technique. “JUNO” is a training respiratory function monitor that provides feedback on mask leak, ventilatory rate, and tidal volume.

Objective

To evaluate whether the use of the JUNO improves face mask ventilation techniques in manikin models.

Methods

We conducted an observational cross‐over study with our unit staff. Following instructions, each participant performed a single‐person technique, followed by a two‐person technique with no JUNO feedback. This was repeated with JUNO feedback visible. A similar sequence was performed both in term and preterm manikins, giving a total of 8 sequences. Each participant was instructed to perform 1 min of positive pressure ventilation providing 30 inflations/minute for all of the simulations. Each of the simulation data underwent a data cleaning process.

Results

Thirty‐eight subjects provided a total of 304 sequences of positive pressure ventilation. A total of 13 354 inflations were analyzed. The feedback group had significantly lower rates of inflations with leak >60%, lower rates of excessive tidal volumes, lower mean leak percentage, and a lower mean inspiratory tidal volumes. When analyzed based on the technique (single person and two‐person), similar positive results were noted in the “feedback group”, across all the strata of staff. All of the staff reported that JUNO improved their ventilation technique and would recommend it for staff training.

Conclusions

The use of JUNO significantly improved mask ventilation consistently across manikin types, staff roles, and techniques (either single‐ or two‐person).

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Cross-Over StudiesManikinsMasksTidal VolumeHumansInfant, NewbornPositive-Pressure RespirationMonitoring, PhysiologicAirway ManagementRespiration, Artificial

Stichwörter der Autor:innen

respiratory function monitorresuscitationtraining

Autor:innen und Institutionen

  • Neonatal Unit, James Cook University Hospital, Middlesbrough, UK.
  • Clinical Academic Office, Faculty of Medical Sciences, Newcastle University, Newcastle upon Tyne, UK.
  • University of Durham, Durham, UK.
  • School of Computing, Engineering and Digital Technologies, Teesside University, Middlesbrough, UK.

ORCID-Profile: Prakash Kannan Loganathan

Zitieren

APA

Loganathan, P. K., Ashton, C., Harrold, E., Wigston, S., Doan, L. M. T., & Occhipinti, A. (2025). Use of real‐time respiratory function monitor improves neonatal face mask ventilation: Cross‐over simulation study. Pediatric Anesthesia, 35(1), 66-74. https://doi.org/10.1111/pan.15020

Vancouver

Loganathan PK, Ashton C, Harrold E, Wigston S, Doan LMT, Occhipinti A. Use of real‐time respiratory function monitor improves neonatal face mask ventilation: Cross‐over simulation study. Pediatric Anesthesia. 2025;35(1):66-74. doi:10.1111/pan.15020

BibTeX
@article{Loganathan2025,
  author = {Loganathan, Prakash Kannan and Ashton, Charlotte and Harrold, Emily and Wigston, Sarah and Doan, Le Minh Thao and Occhipinti, Annalisa},
  title = {Use of real‐time respiratory function monitor improves neonatal face mask ventilation: Cross‐over simulation study},
  journal = {Pediatric Anesthesia},
  year = {2025},
  volume = {35},
  number = {1},
  pages = {66--74},
  doi = {10.1111/pan.15020}
}
Kostenlos starten →