Effect of vertical pump position on start‐up fluid delivery of syringe pumps used for microinfusion
Pediatric Anesthesia · 2023
DOI: 10.1111/pan.14750 ↗Kurz gefasst
Fachartikel in der Zeitschrift Pediatric Anesthesia, 2023. Verfasst von Markus Weiss und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Anesthesia, University Children's Hospital, Zurich, Switzerland. Hauptthemen laut MEDLINE: Infusion Pumps, Central Venous Catheters. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 16 Literaturangaben.
Abstract (Originalsprache)
Background
Connection and opening a syringe infusion pump to a central venous line can lead to acute anterograde or retrograde fluid shifts depending on the level of central venous pressure. This may lead to bolus events or to prolonged lag times of intravenous drug delivery, being particularly relevant when administering vasoactive or inotropic drugs in critically ill patients using microinfusion. The aim of this study was to assess the effect of syringe pump positioning at different vertical heights on start‐up fluid delivery before versus after purging and connection the pump to the central venous catheter.
Methods
This in vitro study measured ante‐ and retrograde infusion volumes delivered to the central venous line after starting the syringe pump at a set infusion rate of 1 mL/h. In setup one, the pump was first positioned to vertical levels of +43 cm or –43 cm and then purged and connected to a central venous catheter. In setup two, the pump was first purged and connected at zero level and secondarily positioned to a vertical level of +43 cm or –43 cm. Central venous pressure was adjusted to 10 mmHg in both setups.
Results
Positioning of the pump prior to purging and connection to the central venous catheter resulted in a better start‐up performance with delivered fluid closer to programmed and expected infusion volumes when compared to the pump first purged, connected, and then positioned. Significant backflow volumes were observed with the pump purged and connected first and then positioned below zero level. No backflow was measured with the pump positioned first below zero level and then purged and connected.
Conclusions
Syringe infusion pump assemblies should be positioned prior to purging and connection to a central venous catheter line when starting a new drug, particularly when administering highly concentrated vasoactive or inotropic drugs delivered at low flow rates.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Anesthesia, University Children's Hospital, Zurich, Switzerland.
- Department of Intensive Care, University Hospital and University of Zurich, Zurich, Switzerland.
- Department of Neonatology, Newborn Research, University Hospital and University of Zurich, Zurich, Switzerland.
- Department of Intensive Care, Kantonsspital, Winterthur, Switzerland.
- Deutsches Herzzentrum der Charité (DHZC) Medicine, Department of Cardiac Anesthesiology and Intensive Care, Berlin, Germany.
- Charité-Universitätsmedizin Berlin, Corporate Member of Freie Universität Berlin and Humboldt-Universität zu Berlin, Berlin, Germany.
ORCID-Profile: Markus Weiss, Pedro David Wendel-Garcia, Beate Grass, Philipp Karl Buehler, Maren Kleine-Brueggeney
Zitieren
APA
Weiss, M., Wendel-Garcia, P. D., Grass, B., Buehler, P. K., & Kleine-Brueggeney, M. (2023). Effect of vertical pump position on start‐up fluid delivery of syringe pumps used for microinfusion. Pediatric Anesthesia. https://doi.org/10.1111/pan.14750
Vancouver
Weiss M, Wendel-Garcia PD, Grass B, Buehler PK, Kleine-Brueggeney M. Effect of vertical pump position on start‐up fluid delivery of syringe pumps used for microinfusion. Pediatric Anesthesia. 2023. doi:10.1111/pan.14750
BibTeX
@article{Weiss2023,
author = {Weiss, Markus and Wendel-Garcia, Pedro David and Grass, Beate and Buehler, Philipp Karl and Kleine-Brueggeney, Maren},
title = {Effect of vertical pump position on start‐up fluid delivery of syringe pumps used for microinfusion},
journal = {Pediatric Anesthesia},
year = {2023},
doi = {10.1111/pan.14750}
}
EUVIMED