EUVIMEDEUVIMED entdecken
Fallbericht

2:1 electronic AV block due to inappropriate automatic post‐ventricular atrial refractory period extension

Pasquale Crea, Paolo Bellocchi, Federica Cocuzza, Antonino Micari iD, Lilia Oreto

Pacing and Clinical Electrophysiology · 2024

DOI: 10.1111/pace.14982 ↗

Kurz gefasst

Fallbericht in der Zeitschrift Pacing and Clinical Electrophysiology, 2024. Verfasst von Pasquale Crea und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Clinical and Experimental Medicine-Cardiology Unit, University of Messina, Messina, Italy. Hauptthemen laut MEDLINE: Atrioventricular Block, Pacemaker, Artificial. Die Arbeit nennt 3 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

A 16‐year‐old female with dual‐chamber pacemaker (Medtronic Azure XT DR), due to symptomatic third‐degree congenital atrioventricular (AV) block, presented to our ambulatory with dizziness and presyncopal episodes preceded by prodromes, occurring over the last few months. The device was programmed in DDD mode with an upper rate of 150 bpm. A head‐up Tilt Test (HUTT) revealed the unexpected emergence of 2:1 electronic AV block at a sinus rate of 130 bpm.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Atrioventricular BlockPacemaker, ArtificialHumansFemaleAdolescentElectrocardiographyTilt-Table Test

Stichwörter der Autor:innen

2:1 AV blockalgorithmautoPVARPdevicepacemaker

Autor:innen und Institutionen

  • Department of Clinical and Experimental Medicine-Cardiology Unit, University of Messina, Messina, Italy.
  • Department of Clinical and Experimental Medicine-Paediatric Cardiology Unit, University of Messina, Messina, Italy.

ORCID-Profile: Antonino Micari

Zitieren

APA

Crea, P., Bellocchi, P., Cocuzza, F., Micari, A., & Oreto, L. (2024). 2:1 electronic AV block due to inappropriate automatic post‐ventricular atrial refractory period extension. Pacing and Clinical Electrophysiology. https://doi.org/10.1111/pace.14982

Vancouver

Crea P, Bellocchi P, Cocuzza F, Micari A, Oreto L. 2:1 electronic AV block due to inappropriate automatic post‐ventricular atrial refractory period extension. Pacing and Clinical Electrophysiology. 2024. doi:10.1111/pace.14982

BibTeX
@article{Crea2024,
  author = {Crea, Pasquale and Bellocchi, Paolo and Cocuzza, Federica and Micari, Antonino and Oreto, Lilia},
  title = {2:1 electronic AV block due to inappropriate automatic post‐ventricular atrial refractory period extension},
  journal = {Pacing and Clinical Electrophysiology},
  year = {2024},
  doi = {10.1111/pace.14982}
}
Kostenlos starten →