Proceed with caution: Standard protocol exercise stress tests fail to replicate the diagnostic utility of supine‐stand tests for long QT syndrome
Pacing and Clinical Electrophysiology · 2024
DOI: 10.1111/pace.14945 ↗Kurz gefasst
Studienprotokoll in der Zeitschrift Pacing and Clinical Electrophysiology, 2024. Verfasst von Alexa M Pinsky und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiovascular Medicine (Division of Heart Rhythm Services), Windland Smith Rice Genetic Heart Rhythm Clinic, Mayo Clinic, Rochester, M…. Hauptthemen laut MEDLINE: Exercise Test, Long QT Syndrome. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 16 Literaturangaben. Gefördert von: NCATS NIH HHS.
Abstract (Originalsprache)
Background
Long QT syndrome (LQTS) is a sudden death predisposing condition characterized by ECG‐derived prolongation of the QT interval. Previous studies have demonstrated that the supine‐stand test may aid in the diagnosis of LQTS as patients fail to shorten their QT interval in response to standing up. The aim of this study was to evaluate the diagnostic accuracy of ECG data derived from standard protocol, clinically performed treadmill exercise stress tests (TESTs) in their ability to mimic the formal supine‐stand test.
Methods
We performed a retrospective review of 478 TESTs from patients evaluated for LQTS. Patients referred for evaluation of LQTS but who were dismissed as normal served as controls. Heart rate & QT values were obtained from standard protocol TESTs.
Results
Overall, 243 patients with LQTS (125 LQT1, 63 LQT2, 55 LQT3; 146 [60%] female, mean age at TEST 30 ± 17 years) and 235 controls (142 [60%] female, mean age 24 ± 15 years) were included. The paired ΔQTc (QTc Stand ‐QTc Supine ) was similar between LQTS (−5 ± 26) and controls (−2 ± 25; p = .2). During position change, the QT interval shortened by ≥20 ms in 33% of LQTS patients, remained unchanged in 62%, and increased in 5% of LQTS patients which was similar to controls (shortened in 40%, unchanged in 54%, and increased in 6% of controls; p = .2). Receiver‐operator curve analysis to test the diagnostic ability of supine‐stand ΔQT performed poorly in differentiating LQTS from controls with an of AUC 0.52 ( p = .4).
Conclusion
TESTs should be used with caution when trying to interpret supine‐stand changes for diagnosis of LQTS.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Cardiovascular Medicine (Division of Heart Rhythm Services), Windland Smith Rice Genetic Heart Rhythm Clinic, Mayo Clinic, Rochester, Minnesota, USA.
- Department of Pediatric and Adolescent Medicine (Division of Pediatric Cardiology), Mayo Clinic, Rochester, Minnesota, USA.
ORCID-Profile: Alexa M Pinsky, J Martijn Bos
Förderung
- NCATS NIH HHS
Zitieren
APA
Pinsky, A. M., Kulkarni, V. K., Bos, J. M., Neves, R., Allison, T. G., & Ackerman, M. J. (2024). Proceed with caution: Standard protocol exercise stress tests fail to replicate the diagnostic utility of supine‐stand tests for long QT syndrome. Pacing and Clinical Electrophysiology. https://doi.org/10.1111/pace.14945
Vancouver
Pinsky AM, Kulkarni VK, Bos JM, Neves R, Allison TG, Ackerman MJ. Proceed with caution: Standard protocol exercise stress tests fail to replicate the diagnostic utility of supine‐stand tests for long QT syndrome. Pacing and Clinical Electrophysiology. 2024. doi:10.1111/pace.14945
BibTeX
@article{Pinsky2024,
author = {Pinsky, Alexa M and Kulkarni, Veda K and Bos, J Martijn and Neves, Raquel and Allison, Thomas G and Ackerman, Michael J},
title = {Proceed with caution: Standard protocol exercise stress tests fail to replicate the diagnostic utility of supine‐stand tests for long QT syndrome},
journal = {Pacing and Clinical Electrophysiology},
year = {2024},
doi = {10.1111/pace.14945}
}
EUVIMED