EUVIMEDEUVIMED entdecken
Randomisierte kontrollierte Studie

Intracardiac echocardiography guided electrified J‐wire trans‐septal puncture: A prospective randomized controlled trial

Hui Cheng iD, Abdul Basit Bangash iD, Yuhang Tao, Ruhong Jiang iD, Qiang Liu, Cheng Cheng, Guosheng Fu iD, Chenyang Jiang iD

Pacing and Clinical Electrophysiology · 2024

DOI: 10.1111/pace.14916 ↗

Kurz gefasst

Randomisierte kontrollierte Studie in der Zeitschrift Pacing and Clinical Electrophysiology, 2024. Verfasst von Hui Cheng und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiology, Sir Run Run Shaw Hospital, School of Medicine, Zhejiang University, Key Laboratory of Cardiovascular Intervention and Regen…. Hauptthemen laut MEDLINE: Atrial Fibrillation, Catheter Ablation. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 17 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Application of electrocautery to a J‐wire is used to perform transseptal puncture (TSP), but with limited evidence supporting safety and efficacy. We conducted a prospective randomized controlled trial to evaluate the safety and efficacy of this technique.

Methods

Two hundred consecutive patients were randomized in a 1:1 fashion to either the ICE‐guided electrified J‐wire TSP group or a conventional Brockenbrough (BRK) needle TSP group. The TSP was performed with a 0.032″ guidewire under 20 W, “coag” mode and was compared to TSP using the BRK needle. The primary safety endpoints were complications related to TSP. The primary efficacy endpoints included the TSP success rate, the total TSP time, and the total procedure time.

Results

All patients complete the procedure safely. The electrified J‐wire TSP group had a significantly shorter TSP time than BRK needle TSP group. The total procedure time, number of TSP attempts required to achieve successful LA access, width of the intra‐atrial shunt at the end of ablation were similar between the two groups. The incidence of new cerebral infarction detected by MRI were similar between the two groups (3/32 patients in the J‐wire TSP group and 2/26 patients in conventional BRK TSP group, p = .82). And no difference in the incidence of residual intra‐atrial shunt (4.3% vs. 6%, p = .654) during the 3‐month's follow up.

Conclusion

Using an electrified J‐wire for TSP under the guidance of ICE appears to be as safe as and more efficient than conventional BRK needle TSP, which may be especially useful in the era of non‐fluoroscopy AF ablation.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Atrial FibrillationCatheter AblationHumansHeart SeptumProspective StudiesPuncturesEchocardiographyTreatment Outcome

Stichwörter der Autor:innen

ablationatrial fibrillationelectrocauteryfossa ovalisintracardiac echocardiographytrans-septal puncture

Autor:innen und Institutionen

  • Department of Cardiology, Sir Run Run Shaw Hospital, School of Medicine, Zhejiang University, Key Laboratory of Cardiovascular Intervention and Regenerative Medicine of Zhejiang Province, Hangzhou, Zhejiang, P. R. China.
  • Department of Cardiology, The Second Hospital of Anhui Medical University, HeFei, Anhui, P. R. China.

ORCID-Profile: Hui Cheng, Abdul Basit Bangash, Ruhong Jiang, Guosheng Fu, Chenyang Jiang

Zitieren

APA

Cheng, H., Bangash, A. B., Tao, Y., Jiang, R., Liu, Q., Cheng, C., Fu, G., & Jiang, C. (2024). Intracardiac echocardiography guided electrified J‐wire trans‐septal puncture: A prospective randomized controlled trial. Pacing and Clinical Electrophysiology. https://doi.org/10.1111/pace.14916

Vancouver

Cheng H, Bangash AB, Tao Y, Jiang R, Liu Q, Cheng C, et al. Intracardiac echocardiography guided electrified J‐wire trans‐septal puncture: A prospective randomized controlled trial. Pacing and Clinical Electrophysiology. 2024. doi:10.1111/pace.14916

BibTeX
@article{Cheng2024,
  author = {Cheng, Hui and Bangash, Abdul Basit and Tao, Yuhang and Jiang, Ruhong and Liu, Qiang and Cheng, Cheng and Fu, Guosheng and Jiang, Chenyang},
  title = {Intracardiac echocardiography guided electrified J‐wire trans‐septal puncture: A prospective randomized controlled trial},
  journal = {Pacing and Clinical Electrophysiology},
  year = {2024},
  doi = {10.1111/pace.14916}
}
Kostenlos starten →