Remote monitoring of implantable cardiac monitors in patients with unexplained syncope: Predictors of false‐positive alert episodes
Pacing and Clinical Electrophysiology · 2023
DOI: 10.1111/pace.14851 ↗Kurz gefasst
Fachartikel in der Zeitschrift Pacing and Clinical Electrophysiology, 2023. Verfasst von Vincenzo Russo und 11 weiteren Autor:innen. Beteiligte Einrichtung: Cardiology and Syncope Unit, Department of Medical Translational Sciences, University of Campania "Luigi Vanvitelli" - Monaldi Hospital, Naples, Ital…. Hauptthemen laut MEDLINE: Arrhythmias, Cardiac, Electrocardiography. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 16 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0).
Abstract (Originalsprache)
Background
Remote monitoring is recommended for patients with implantable cardiac monitors (ICMs), but compared to other cardiac implantable devices, ICMs are less accurate and transmit a higher number of alerts.
Objective
The aim of this study was to investigate the predictors of false‐positive (FP) arrhythmic alerts in patients with unexplained syncope who were implanted with ICM and followed by an automatic remote monitoring system.
Methods
We retrospectively evaluated all consecutive patients who received a long‐sensing vector ICM for unexplained syncope between January 2019 to September 2021 at our Syncope Unit. The primary endpoint was the incidence of the first FP episode. The secondary endpoints included assessing the incidence of FP episodes for all types of algorhythms and indentifying the reasons for the misdetection of these episodes.
Results
Among 105 patients (44.8% males, median age 51 years), 51 (48.6%) transmitted at least one FP alert during a median follow‐up of 301 days. The presence of pre‐ventricular complexes (PVCs) on the resting electrocardiogram was the only clinical characteristic associated with an increased risk of FP alerts (adjusted Hazard ratio [HR] 5.76 [2.66–12.4], p = 0.010). The other significant device‐related variables were a low‐frequency filter at 0.05 Hz versus the default 0.5 Hz (adjusted HR 3.82 [1.38–10.5], p = 0.010) and the R‐wave amplitude (adjusted HR 0.35 [0.13–0.99], p = 0.049).
Conclusion
Patients who have PVCs are at higher risk of inappropriate ICM activations. To reduce the occurrence of FP alerts, it may be beneficial to target a large R‐wave amplitude during device insertion and avoid programming a low‐frequency filter at 0.05 Hz.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Cardiology and Syncope Unit, Department of Medical Translational Sciences, University of Campania "Luigi Vanvitelli" - Monaldi Hospital, Naples, Italy.
- Clinical Research Unit, Biotronik Italia S.p.A., Milan, Italy.
ORCID-Profile: Vincenzo Russo
Zitieren
APA
Russo, V., Covino, S., De Pasquale, V., Parente, E., Comune, A., Rago, A., Papa, A. A., Ammendola, E., Spadaro Guerra, A., Napoli, P., Golino, P., & Nigro, G. (2023). Remote monitoring of implantable cardiac monitors in patients with unexplained syncope: Predictors of false‐positive alert episodes. Pacing and Clinical Electrophysiology. https://doi.org/10.1111/pace.14851
Vancouver
Russo V, Covino S, De Pasquale V, Parente E, Comune A, Rago A, et al. Remote monitoring of implantable cardiac monitors in patients with unexplained syncope: Predictors of false‐positive alert episodes. Pacing and Clinical Electrophysiology. 2023. doi:10.1111/pace.14851
BibTeX
@article{Russo2023,
author = {Russo, Vincenzo and Covino, Simona and De Pasquale, Valentina and Parente, Erika and Comune, Angelo and Rago, Anna and Papa, Andrea Antonio and Ammendola, Ernesto and Spadaro Guerra, Andrea and Napoli, Paola and Golino, Paolo and Nigro, Gerardo},
title = {Remote monitoring of implantable cardiac monitors in patients with unexplained syncope: Predictors of false‐positive alert episodes},
journal = {Pacing and Clinical Electrophysiology},
year = {2023},
doi = {10.1111/pace.14851}
}
EUVIMED