EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Finite Element Analysis of Six Internal Fixations in the Treatment of Pauwels Type III Femoral Neck Fracture

Xiang Sun, Zhe Han iD, Dongdong Cao, Chao Han, Mengqi Xie, Xiantie Zeng, Qiang Dong

Orthopaedic Surgery · 2024-07 · 16 (7) · 1695-1709

DOI: 10.1111/os.14069 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Orthopaedic Surgery, online erschienen am 15.05.2024. Verfasst von Xiang Sun und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Hip Trauma, Tianjin Hospital, Tianjin, China. Hauptthemen laut MEDLINE: Fracture Fixation, Internal, Finite Element Analysis, Femoral Neck Fractures, Bone Screws. Die Arbeit wurde bisher 16-mal zitiert (Crossref) und nennt 53 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: The Science and Technology Project of Tianjin Health Commission.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objective

The current investigation sought to utilize finite element analysis to replicate the biomechanical effects of different fixation methods, with the objective of establishing a theoretical framework for the optimal choice of modalities in managing Pauwels type III femoral neck fractures.

Methods

The Pauwels type III fracture configuration, characterized by angles of 70°, was simulated in conjunction with six distinct internal fixation methods, including cannulated compression screw (CCS), dynamic hip screw (DHS), DHS with de‐rotational screw (DS), CCS with medial buttress plate (MBP), proximal femoral nail anti‐rotation (PFNA), and femoral neck system (FNS). These models were developed and refined using Geomagic and SolidWorks software. Subsequently, finite element analysis was conducted utilizing Ansys software, incorporating axial loading, torsional loading, yield loading and cyclic loading.

Results

Under axial loading conditions, the peak stress values for internal fixation and the femur were found to be highest for CCS (454.4; 215.4 MPa) and CCS + MBP (797.2; 284.2 MPa), respectively. The corresponding maximum and minimum displacements for internal fixation were recorded as 6.65 mm for CCS and 6.44 mm for CCS + MBP. When subjected to torsional loading, the peak stress values for internal fixation were highest for CCS + MBP (153.6 MPa) and DHS + DS (72.8 MPa), while for the femur, the maximum and minimum peak stress values were observed for CCS + MBP (119.3 MPa) and FNS (17.6 MPa), respectively. Furthermore, the maximum and minimum displacements for internal fixation were measured as 0.249 mm for CCS + MBP and 0.205 mm for PFNA. Additionally, all six internal fixation models showed excellent performance in terms of yield load and fatigue life.

Conclusion

CCS + MBP had the best initial mechanical stability in treatment for Pauwels type III fracture. However, the MBP was found to be more susceptible to shear stress, potentially increasing the risk of plate breakage. Furthermore, the DHS + DS exhibited superior biomechanical stability compared to CCS, DHS, and PFNA, thereby offering a more conducive environment for fracture healing. Additionally, it appeared that FNS represented a promising treatment strategy, warranting further validation in future studies.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Fracture Fixation, InternalFinite Element AnalysisFemoral Neck FracturesBone ScrewsBone PlatesHumansBiomechanical PhenomenaWeight-BearingBone NailsStress, Mechanical

Stichwörter der Autor:innen

Biomechanical studyFinite element analysisInternal fixationPauwels type III femoral neck fracture

Autor:innen und Institutionen

  • Department of Hip Trauma, Tianjin Hospital, Tianjin, China.
  • Tianjin University of Traditional Chinese Medicine, Tianjin, China.
  • Department of Foot and Ankle Surgery, Tianjin Hospital, Tianjin, China.

ORCID-Profile: Zhe Han

Förderung

  • The Science and Technology Project of Tianjin Health Commission

Zitieren

APA

Sun, X., Han, Z., Cao, D., Han, C., Xie, M., Zeng, X., & Dong, Q. (2024). Finite Element Analysis of Six Internal Fixations in the Treatment of Pauwels Type III Femoral Neck Fracture. Orthopaedic Surgery, 16(7), 1695-1709. https://doi.org/10.1111/os.14069

Vancouver

Sun X, Han Z, Cao D, Han C, Xie M, Zeng X, et al. Finite Element Analysis of Six Internal Fixations in the Treatment of Pauwels Type III Femoral Neck Fracture. Orthopaedic Surgery. 2024;16(7):1695-1709. doi:10.1111/os.14069

BibTeX
@article{Sun2024,
  author = {Sun, Xiang and Han, Zhe and Cao, Dongdong and Han, Chao and Xie, Mengqi and Zeng, Xiantie and Dong, Qiang},
  title = {Finite Element Analysis of Six Internal Fixations in the Treatment of Pauwels Type III Femoral Neck Fracture},
  journal = {Orthopaedic Surgery},
  year = {2024},
  volume = {16},
  number = {7},
  pages = {1695--1709},
  doi = {10.1111/os.14069}
}
Kostenlos starten →