EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Radiometric and safety assessment of a ‘low‐level red‐light’ myopia control device

Josh Richards iD, Jennifer J Hunter, Javier Gantes-Nuñez iD, Arthur Bradley, Pete Kollbaum iD

Ophthalmic and Physiological Optics · 2025

DOI: 10.1111/opo.70009 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Ophthalmic and Physiological Optics, 2025. Verfasst von Josh Richards und 4 weiteren Autor:innen. Beteiligte Einrichtung: School of Optometry, Indiana University, Bloomington, Indiana, USA. Hauptthemen laut MEDLINE: Myopia, Radiometry, Phototherapy, Refraction, Ocular. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 40 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: CooperVision.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

Recent literature has demonstrated that ‘low‐level red‐light therapy’ may be effective at slowing axial elongation in children, but reports have questioned the safety of some red‐light devices. The current work explored the beam characteristics and hazard potential for a myopia control red‐light device.

Methods

The optical design and exiting beam characteristics of a sample red‐light myopia control device (SECONEE sky‐n1201a) were quantified by measuring beam divergence and total flux passing through artificial pupils of 1–9 mm diameter placed at the corneal plane. Retinal exposure levels were compared to American National Standards Institute (ANSI) limits for ophthalmic instruments (Z80.36‐2021).

Results

Two 655 nm laser diodes created exiting red‐light beams (~9 mm in diameter) that were approximately collimated by an internal pair of +10.00 D eye‐piece lenses resulting in exit vergences ranging from −3.25 to −7.75 D as interpupillary distance settings were adjusted from 52 to 70 mm. Radiant power (mW) passing through the artificial pupils increased from 0.005 to 0.65 mW as pupil diameters increased from 1 to 9 mm. Using ANSI Z80.36‐2021, time‐averaged retinal radiant exposures (J/cm 2 ) exceeded the thermal and photochemical limits for pupil diameters >1 mm and >5 mm, respectively. The time to reach potential photochemical radiation hazard ( t max ) was less than 100 s for pupil diameters >5 mm.

Conclusions

The optical design and resulting beam characteristics of this sample red‐light myopia control device generate retinal exposure levels that vary with pupil diameter, accommodation and eye movements. Conservative estimates reveal retinal radiant exposures that can exceed ANSI safety limits. The ‘low‐level’ moniker is inappropriate for a device that either exceeds or is just below the ANSI threshold for potential retinal hazard.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

MyopiaRadiometryPhototherapyRefraction, OcularHumansEquipment DesignRetina

Stichwörter der Autor:innen

low‐level red‐light therapymyopia controlphotobiomodulation therapyretinal hazard

Autor:innen und Institutionen

  • School of Optometry, Indiana University, Bloomington, Indiana, USA.
  • School of Optometry and Vision Science, University of Waterloo, Waterloo, Ontario, Canada.
  • Midwestern University, Arizona College of Optometry, Glendale, Arizona, USA.
  • CooperVision, Inc., San Ramon, California, USA.

ORCID-Profile: Josh Richards, Javier Gantes-Nuñez, Pete Kollbaum

Förderung

  • CooperVision

Zitieren

APA

Richards, J., Hunter, J. J., Gantes-Nuñez, J., Bradley, A., & Kollbaum, P. (2025). Radiometric and safety assessment of a ‘low‐level red‐light’ myopia control device. Ophthalmic and Physiological Optics. https://doi.org/10.1111/opo.70009

Vancouver

Richards J, Hunter JJ, Gantes-Nuñez J, Bradley A, Kollbaum P. Radiometric and safety assessment of a ‘low‐level red‐light’ myopia control device. Ophthalmic and Physiological Optics. 2025. doi:10.1111/opo.70009

BibTeX
@article{Richards2025,
  author = {Richards, Josh and Hunter, Jennifer J and Gantes-Nuñez, Javier and Bradley, Arthur and Kollbaum, Pete},
  title = {Radiometric and safety assessment of a ‘low‐level red‐light’ myopia control device},
  journal = {Ophthalmic and Physiological Optics},
  year = {2025},
  doi = {10.1111/opo.70009}
}
Kostenlos starten →