A novel wearable sensor for objective measurement of distance and illumination
Ophthalmic and Physiological Optics · 2025-07 · 45 (5) · 1067-1079
DOI: 10.1111/opo.13523 ↗Kurz gefasst
Fachartikel in der Zeitschrift Ophthalmic and Physiological Optics, online erschienen am 03.05.2025. Verfasst von Raman Prasad Sah und 2 weiteren Autor:innen. Beteiligte Einrichtung: University of Houston College of Optometry, Houston, Texas, USA. Hauptthemen laut MEDLINE: Wearable Electronic Devices, Lighting, Eyeglasses, Myopia. Die Arbeit wurde bisher 11-mal zitiert (Crossref) und nennt 47 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: NEI NIH HHS.
Abstract (Originalsprache)
Purpose
To evaluate the rangefinding and light sensing capabilities of a novel wearable sensor, the Visual Environment Evaluation Tool (VEET) developed for myopia research.
Methods
The VEET 1.0 (Meta, LLC) is a temple‐integrated system mounted on a spectacle frame. Both the left and right temples house four sensors each, angled straight ahead and 20° downward/4° nasal, respectively. For validation, VEET‐mounted spectacles were placed on a mannequin head. An infrared camera was used to capture the spatial characteristics of the rangefinding beam. Distance measurements were collected against a wall for 5–400 cm. The accuracy of distance measurements for different target types, sizes and tilt was assessed. Ambient illumination was captured in different indoor and outdoor settings and compared to a lux meter.
Results
All devices ( N = 20) were capable of capturing distances for the full range evaluated, 5–400 cm, against a wall. There was a strong relationship between the actual and measured distances, with a slope of 1.01 ± 0.003 and 0.95 ± 0.007, for the left and right temples, respectively. Distance measurements were repeatable across different target types, including paper and tablet. Mean beam diameter of five tested devices was 52.2° ± 7.5°. The VEET effectively measured distances across different target sizes, ranging from 2 × 2 cm and larger and target tilt ±60°. Illumination measurements across different indoor and outdoor settings demonstrated a strong linear relationship with lux meter readings ( R 2 = 0.99 and 0.78 for the left and right temples, respectively), effectively distinguishing indoor (
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- University of Houston College of Optometry, Houston, Texas, USA.
ORCID-Profile: Raman Prasad Sah, Lisa A Ostrin
Förderung
- NEI NIH HHS
Zitieren
APA
Sah, R. P., Narra, P. K., & Ostrin, L. A. (2025). A novel wearable sensor for objective measurement of distance and illumination. Ophthalmic and Physiological Optics, 45(5), 1067-1079. https://doi.org/10.1111/opo.13523
Vancouver
Sah RP, Narra PK, Ostrin LA. A novel wearable sensor for objective measurement of distance and illumination. Ophthalmic and Physiological Optics. 2025;45(5):1067-1079. doi:10.1111/opo.13523
BibTeX
@article{Sah2025,
author = {Sah, Raman Prasad and Narra, Pavan Kalyan and Ostrin, Lisa A},
title = {A novel wearable sensor for objective measurement of distance and illumination},
journal = {Ophthalmic and Physiological Optics},
year = {2025},
volume = {45},
number = {5},
pages = {1067--1079},
doi = {10.1111/opo.13523}
}
EUVIMED