EUVIMEDEUVIMED entdecken
Open Access · CC BY-NC-ND 4.0

Influence of brightness artefacts on corneal densitometry

Ana R Arizcuren iD, Laura Remón iD, Alejandra Consejo iD

Ophthalmic and Physiological Optics · 2025-05 · 45 (3) · 637-643

DOI: 10.1111/opo.13483 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Ophthalmic and Physiological Optics, online erschienen am 05.03.2025. Verfasst von Ana R Arizcuren und 2 weiteren Autor:innen. Beteiligte Einrichtung: Aragon Institute for Engineering Research (I3A), University of Zaragoza, Zaragoza, Spain. Hauptthemen laut MEDLINE: Densitometry, Artifacts, Cornea. Die Arbeit wurde bisher 8-mal zitiert (Crossref) und nennt 22 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: MCIN/AEI/10.13039/501100011033 and European Union "NextGenerationEU"/PRTR.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

To quantify objectively the influence of brightness artefacts inherent in Scheimpflug tomography on corneal densitometry (CD) estimates.

Methods

Fifty‐seven healthy participants aged 30.9 ± 13.7 years (range 17–60 years) were examined with a Ziemer Galilei G2 Scheimpflug tomographer. Images were automatically segmented to isolate the cornea, iris, crystalline lens and corneoscleral lateral brightness regions. CD was calculated as the corneal mean pixel intensity (MPI). Statistical analysis explored the relationship between brightness artefacts and CD across different corneal layers, while also considering age and biometric parameters such as anterior chamber depth and pupil size.

Results

Brightness artefacts accounted for 6.56 ± 1.93% of the pixels in the images, with lateral brightness being the largest source of noise (4.19 ± 0.97%). A strong positive correlation was found between CD and overall brightness artefacts ( r = 0.57, p < 0.001). Age was also significantly correlated with both CD ( r = 0.65, p < 0.001) and brightness ( r = 0.29, p = 0.01). Mediation analysis revealed that age accounts for 60.0% of CD variability, while brightness artefacts contributed 40.0%. Brightness artefacts had the strongest effect on the corneal periphery beyond 7.5 mm as well as the stromal layer.

Conclusion

Brightness artefacts significantly influence CD estimates, particularly in the peripheral cornea. Accurate quantification of these artefacts is essential for proper clinical assessment of corneal transparency, especially in older populations. Future research should focus on identifying the specific sources of artefacts to refine CD measurement techniques.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

DensitometryArtifactsCorneaHumansMaleFemaleAdolescentYoung AdultAdultMiddle AgedTomographyLight

Stichwörter der Autor:innen

Scheimpflug imagingbrightness artefactscorneal densitometrycorneal transparencyimage processing

Autor:innen und Institutionen

  • Aragon Institute for Engineering Research (I3A), University of Zaragoza, Zaragoza, Spain.
  • Department of Applied Physics, University of Zaragoza, Zaragoza, Spain.

ORCID-Profile: Ana R Arizcuren, Laura Remón, Alejandra Consejo

Förderung

  • MCIN/AEI/10.13039/501100011033 and European Union "NextGenerationEU"/PRTR

Zitieren

APA

Arizcuren, A. R., Remón, L., & Consejo, A. (2025). Influence of brightness artefacts on corneal densitometry. Ophthalmic and Physiological Optics, 45(3), 637-643. https://doi.org/10.1111/opo.13483

Vancouver

Arizcuren AR, Remón L, Consejo A. Influence of brightness artefacts on corneal densitometry. Ophthalmic and Physiological Optics. 2025;45(3):637-643. doi:10.1111/opo.13483

BibTeX
@article{Arizcuren2025,
  author = {Arizcuren, Ana R and Remón, Laura and Consejo, Alejandra},
  title = {Influence of brightness artefacts on corneal densitometry},
  journal = {Ophthalmic and Physiological Optics},
  year = {2025},
  volume = {45},
  number = {3},
  pages = {637--643},
  doi = {10.1111/opo.13483}
}
Kostenlos starten →