Fachartikel in der Zeitschrift Ophthalmic and Physiological Optics, 2024. Verfasst von Janneau L J Claessens und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Ophthalmology, University Medical Center Utrecht, Utrecht, the Netherlands. Hauptthemen laut MEDLINE: Visual Acuity, Myopia, Internet, Refraction, Ocular. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 31 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Introduction

Demands for myopia management are rising. A web‐based tool that allows home‐performed self‐assessments of visual acuity (VA) and refractive error may enable hybrid care pathways and aid in identifying those with deteriorating visual performance. The tool has been validated in adult populations, but has yet to be evaluated in children. This study compared home‐performed VA and refraction self‐assessments to conventional measurements obtained at the clinic in a population of myopic children.

Methods

Myopic children aged ≥6 years old were invited to perform web‐based eye tests at home, assisted by a parent. At two myopia control clinics, they also underwent measurements of VA using a Snellen chart and refractive error using cycloplegic autorefraction. Agreement between the tests, repeatability of the web‐based test and associations between clinical characteristics and web‐based test accuracy were evaluated.

Results

A total of 147 children were enrolled, of whom 116 (51% male; mean age 13 ± 3 years; mean spherical equivalent refraction (SEQ) −5.58 ± 3.05) performed the web‐based tests at home. Overall, the home‐performed VA self‐assessment and the Snellen chart assessment at the clinic agreed well (mean difference 0.03 ± 0.11 logMAR). A significant proportional bias was identified (β 0.65, p < 0.001), indicating underestimated web‐based VA scores when the child's vision declined. The sensitivity to detect VA poorer than 0.10 logMAR was 94%; the specificity was 71%. The web‐based refractive error algorithm measured more myopia progression compared to clinic observations (mean difference SEQ 0.40 ± 0.51 dioptres). Age, sex or use of atropine drops were not significantly associated with test accuracy.

Conclusions

The web‐based test for self‐assessing vision, performed at home by children with assistance from their parents, yielded VA scores with a precision similar to Snellen chart testing conducted in a clinical setting. However, the web‐based refractive error algorithm overestimated myopia progression and requires recalibration for this specific age group.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Visual AcuityMyopiaInternetRefraction, OcularVision TestsHumansMaleFemaleChildAdolescentReproducibility of ResultsSelf-Assessment

Stichwörter der Autor:innen

myopiarefractiontelehealthtelemedicinetelemonitoringvisual acuity

Autor:innen und Institutionen

  • Department of Ophthalmology, University Medical Center Utrecht, Utrecht, the Netherlands.
  • Department of Ophthalmology, Erasmus University Medical Center, Rotterdam, the Netherlands.
  • Department of Optometry & Orthoptics, Institute of Allied Health Studies, University of Applied Sciences, Utrecht, the Netherlands.
  • Department of Ophthalmology, Radboud University Medical Center, Nijmegen, the Netherlands.
  • Department of Epidemiology, Erasmus Medical Center, Rotterdam, Netherlands.
  • Institute of Molecular and Clinical Ophthalmology, University of Basel, Basel, Switzerland.

ORCID-Profile: Janneau L J Claessens, Robert P L Wisse

Zitieren

APA

Claessens, J. L. J., Janssen, M. A., van der Zee, C., Polling, J. R., Meester-Smoor, M. A., Klaver, C. C. W., & Wisse, R. P. L. (2024). Evaluating a web‐based visual acuity and refractive error self‐assessment tool in myopic children. Ophthalmic and Physiological Optics. https://doi.org/10.1111/opo.13370

Vancouver

Claessens JLJ, Janssen MA, van der Zee C, Polling JR, Meester-Smoor MA, Klaver CCW, et al. Evaluating a web‐based visual acuity and refractive error self‐assessment tool in myopic children. Ophthalmic and Physiological Optics. 2024. doi:10.1111/opo.13370