Red light instruments for myopia exceed safety limits
Ophthalmic and Physiological Optics · 2024-03 · 44 (2) · 241-248
DOI: 10.1111/opo.13272 ↗Kurz gefasst
Fachartikel in der Zeitschrift Ophthalmic and Physiological Optics, online erschienen am 05.01.2024. Verfasst von Lisa A Ostrin und 1 weiteren Autor:innen. Beteiligte Einrichtung: University of Houston College of Optometry, Houston, Texas, USA. Hauptthemen laut MEDLINE: Red Light, Myopia. Die Arbeit wurde bisher 77-mal zitiert (Crossref) und nennt 41 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: NEI NIH HHS, National Eye Institute.
Abstract (Originalsprache)
Purpose
Low‐level red light (LLRL) therapy has recently emerged as a myopia treatment in children, with several studies reporting significant reduction in axial elongation and myopia progression. The goal of this study was to characterise the output and determine the thermal and photochemical maximum permissible exposure (MPE) of LLRL devices for myopia control.
Methods
Two LLRL devices, a Sky‐n1201a and a Future Vision, were examined. Optical power measurements were made using an integrating sphere radiometer through a 7‐mm diameter aperture, in accordance with ANSI Z136.1‐2014, sections 3.2.3–3.2.4. Retinal spot sizes of the devices were obtained using a model eye and high‐resolution beam profiler. Corneal irradiance, retinal irradiance and MPE were calculated for an eye positioned at the oculars of each device.
Results
Both devices were confirmed to be Class 1 laser products. Findings showed that the Sky‐n1201a delivers laser light as a point source with a 654‐nm wavelength, 0.2 mW power (Ø 7 mm aperture, 10‐cm distance), 1.17 mW/cm 2 corneal irradiance and 7.2 W/cm 2 retinal irradiance (Ø 2 mm pupil). The MPE for photochemical damage is 0.55–7.0 s for 2–7 mm pupils and for thermal damage is 0.41–10 s for 4.25–7 mm pupils. Future Vision delivers the laser as an extended source subtending 0.75 × 0.325°. It has a 652‐nm wavelength, 0.06 mW power (Ø 7 mm aperture, 10 cm distance), 0.624 mW/cm 2 corneal irradiance and 0.08 W/cm 2 retinal irradiance (Ø 2 mm pupil). MPE for photochemical damage is 50–625 s for 2–7 mm pupils.
Discussion
For both of the LLRL devices evaluated here, 3 min of continuous viewing approached or surpassed the MPE, putting the retina at risk of photochemical and thermal damage. Clinicians should be cautious with the use of LLRL therapy for myopia in children until safety standards can be confirmed.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- University of Houston College of Optometry, Houston, Texas, USA.
ORCID-Profile: Lisa A Ostrin
Förderung
- NEI NIH HHS
- National Eye Institute
Zitieren
APA
Ostrin, L. A., & Schill, A. W. (2024). Red light instruments for myopia exceed safety limits. Ophthalmic and Physiological Optics, 44(2), 241-248. https://doi.org/10.1111/opo.13272
Vancouver
Ostrin LA, Schill AW. Red light instruments for myopia exceed safety limits. Ophthalmic and Physiological Optics. 2024;44(2):241-248. doi:10.1111/opo.13272
BibTeX
@article{Ostrin2024,
author = {Ostrin, Lisa A and Schill, Alexander W},
title = {Red light instruments for myopia exceed safety limits},
journal = {Ophthalmic and Physiological Optics},
year = {2024},
volume = {44},
number = {2},
pages = {241--248},
doi = {10.1111/opo.13272}
}
EUVIMED