Repeatability and agreement of central corneal thickness measurements with a new handheld non‐contact pachymeter

John G Lawrenson iD, Simranjit Gill, Isra Masuid, Fardip Rashid

Ophthalmic and Physiological Optics · 2023

DOI: 10.1111/opo.13199 ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift Ophthalmic and Physiological Optics, 2023. Verfasst von John G Lawrenson und 3 weiteren Autor:innen. Beteiligte Einrichtung: Centre for Applied Vision Research, School of Health and Psychological Sciences, City, University of London, London, UK. Hauptthemen laut MEDLINE: Cornea, Corneal Pachymetry. Die Arbeit nennt 33 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Occuity Ltd.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

To compare the repeatability of central corneal thickness (CCT) measurements taken with a new handheld pachymeter (Occuity PM1 pachymeter) and to assess its agreement with ultrasound biometry and two commercially available optical biometers in participants with normal eyes.

Methods

Three consecutive CCT measurements of the right eye of 105 participants with normal corneas were acquired by the PM1 pachymeter, Lenstar LS 900 and Oculus Pentacam HR in a random order. This was followed by three measurements with a handheld ultrasound pachymeter (UP) (Pachmate 2). Repeatability and the repeatability limit were calculated for each device and Bland–Altman limits of agreement (LoA) were determined for the PM1 pachymeter compared to the other devices.

Results

The mean CCT (±SD) was 551.04 ± 33.43, 558.62 ± 31.46, 549.41 ± 31.00 and 539.73 ± 29.50 μm for the PM1 pachymeter, UP, Lenstar and Pentacam, respectively. The repeatability limits (expressed as the within subject SD for repeat measurements) were 14.02, 13.68, 4.99 and 9.90 μm, respectively. The closest agreement was between the PM1 and Lenstar (mean difference = −1.63 μm with LoA 10.72 μm below and 13.97 μm above the readings obtained with the Lenstar). The PM1 underestimated CCT compared to UP (mean difference = 7.58 μm, LoA 24.63 μm below and 9.47 μm above UP). The agreement was lowest between the PM1 and Pentacam (mean difference = −11.30 μm, LoA between 4.29 and 26.89 μm).

Conclusions

The PM1 pachymeter shows excellent precision for CCT measurements across a range of corneal thicknesses in normal eyes and provides a safe and easy‐to‐use alternative to ultrasound pachymetry.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

CorneaCorneal PachymetryHumansReproducibility of ResultsMaleFemaleAdultMiddle AgedYoung AdultBiometryAgedHealthy Volunteers

Stichwörter der Autor:innen

agreementcentral corneal thicknesspachymetryprecision

Autor:innen und Institutionen

  • Centre for Applied Vision Research, School of Health and Psychological Sciences, City, University of London, London, UK.
  • Moorfields Eye Hospital, NHS Foundation Trust, London, UK.
  • Royal Berkshire, NHS Foundation Trust, Reading, UK.

ORCID-Profile: John G Lawrenson

Förderung

  • Occuity Ltd

Zitieren

APA

Lawrenson, J. G., Gill, S., Masuid, I., & Rashid, F. (2023). Repeatability and agreement of central corneal thickness measurements with a new handheld non‐contact pachymeter. Ophthalmic and Physiological Optics. https://doi.org/10.1111/opo.13199

Vancouver

Lawrenson JG, Gill S, Masuid I, Rashid F. Repeatability and agreement of central corneal thickness measurements with a new handheld non‐contact pachymeter. Ophthalmic and Physiological Optics. 2023. doi:10.1111/opo.13199

BibTeX
@article{Lawrenson2023,
  author = {Lawrenson, John G and Gill, Simranjit and Masuid, Isra and Rashid, Fardip},
  title = {Repeatability and agreement of central corneal thickness measurements with a new handheld non‐contact pachymeter},
  journal = {Ophthalmic and Physiological Optics},
  year = {2023},
  doi = {10.1111/opo.13199}
}
Kostenlos starten →