Fachartikel in der Zeitschrift Neurogastroenterology & Motility, online erschienen am 07.10.2026. Verfasst von Aidan D Smires und 7 weiteren Autor:innen. Beteiligte Einrichtung: Kenneth C. Griffin Esophageal Center of Northwestern Medicine, Department of Medicine, Division of Gastroenterology and Hepatology, Feinberg School o…. Hauptthemen laut MEDLINE: Manometry, Esophagogastric Junction, Esophageal Motility Disorders. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 16 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NIDDK NIH HHS, Public Health service (NIDDK.
Abstract (Originalsprache)
Background
Esophagogastric junction outflow obstruction (EGJOO) identified on high‐resolution manometry (HRM) remains a clinically heterogeneous and inconclusive finding. Additional complementary testing with functional lumen imaging probe (FLIP) or timed barium esophagram (TBE) is recommended to confirm clinically relevant obstruction. We aimed to assess the association of novel Four‐Dimensional (4D) HRM metrics against FLIP and TBE outcomes among symptomatic patients with inconclusive EGJOO on HRM.
Methods
Adult patients with inconclusive HRM‐EGJOO who completed corresponding FLIP and TBE studies were identified for inclusion in a cross‐sectional, retrospective study. Primary 4D HRM metrics of EGJ opening, EGJ distensibility index (4D EGJ‐DI), and 4D maximum EGJ diameter were analyzed relative to reference outcomes defined by FLIP and esophagram.
Results
Among 123 patients (mean age 59 years; 60% female), 4D EGJ‐DI and 4D maximum EGJ diameter differed significantly across FLIP and TBE outcome categories (all p < 0.01). Lower 4D EGJ‐DI and lower 4D maximum EGJ diameter were associated with reduced EGJ opening on FLIP and abnormal retention on esophagram. There were 0 patients with normal EGJ opening on FLIP that had 4D maximum EGJ diameter < 8.0 mm and 0 patients with reduced EGJ opening on FLIP that had a 4D EGJ‐DI > 2.2 mm 2 /mmHg and 4D maximum EGJ diameter > 10.2 mm.
Conclusions
By integrating biomechanical assessment of EGJ opening with conventional pressure metrics, 4D HRM provides complementary physiologic information for evaluating esophageal motor function. This additional information may improve characterization of patients with inconclusive EGJOO.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Kenneth C. Griffin Esophageal Center of Northwestern Medicine, Department of Medicine, Division of Gastroenterology and Hepatology, Feinberg School of Medicine, Northwestern University, Chicago, Illinois, USA.
- Center of Excellence in Neurogastroenterology and Motility, Division of Gastroenterology, Department of Medicine, Faculty of Medicine, Chulalongkorn University, Bangkok, Thailand.
- Division of Thoracic Surgery, Department of Surgery, Université de Montréal, Montreal, Quebec, Canada.
- Department of Mechanical Engineering, McCormick School of Engineering, Northwestern University, Evanston, Illinois, USA.
ORCID-Profile: Aidan D Smires, Wenjun Kou, Panyavee Pitisuttithum, Eric Goudie, Muhammed M Alikhan, Neelesh A Patankar, Dustin A Carlson
Förderung
- NIDDK NIH HHS
- Public Health service (NIDDK
Zitieren
APA
Smires, A. D., Kou, W., Pandolfino, J. E., Pitisuttithum, P., Goudie, E., Alikhan, M. M., Patankar, N. A., & Carlson, D. A. (2026). Biomechanical Assessment During High‐Resolution Impedance Manometry Enhances Characterization of Inconclusive Esophagogastric Junction Outflow Obstruction. Neurogastroenterology & Motility, 38(10). https://doi.org/10.1111/nmo.70450
Vancouver
Smires AD, Kou W, Pandolfino JE, Pitisuttithum P, Goudie E, Alikhan MM, et al. Biomechanical Assessment During High‐Resolution Impedance Manometry Enhances Characterization of Inconclusive Esophagogastric Junction Outflow Obstruction. Neurogastroenterology & Motility. 2026;38(10). doi:10.1111/nmo.70450
BibTeX
@article{Smires2026,
author = {Smires, Aidan D and Kou, Wenjun and Pandolfino, John E and Pitisuttithum, Panyavee and Goudie, Eric and Alikhan, Muhammed M and Patankar, Neelesh A and Carlson, Dustin A},
title = {Biomechanical Assessment During High‐Resolution Impedance Manometry Enhances Characterization of Inconclusive Esophagogastric Junction Outflow Obstruction},
journal = {Neurogastroenterology & Motility},
year = {2026},
volume = {38},
number = {10},
doi = {10.1111/nmo.70450}
}
EUVIMED