Fachartikel in der Zeitschrift Neurogastroenterology & Motility, online erschienen am 04.01.2023. Verfasst von Sarvee Moosavi und 7 weiteren Autor:innen. Beteiligte Einrichtung: Division of Gastroenterology, University of British Columbia, Vancouver General Hospital, Vancouver, British Columbia, Canada. Hauptthemen laut MEDLINE: Eosinophilic Esophagitis. Die Arbeit wurde bisher 27-mal zitiert (Crossref) und nennt 17 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: NIDDK NIH HHS, U.S. Public Health Service.
Abstract (Originalsprache)
Background
Eosinophilic esophagitis (EoE) is associated with fibrostenotic remodeling that can be objectively assessed using the functional lumen imaging probe (FLIP). This is typically done using a metric called distensibility plateau (DP). We aimed to describe a novel measure of compliance of the esophageal body and evaluate the associated clinical characteristics in EoE.
Methods
One hundred seventy‐one adult patients with EoE (mean (SD) age 38 (12) years), 31% female and 35 healthy, asymptomatic controls who completed 16‐cm functional luminal imaging probe (FLIP) during endoscopy, were evaluated in a cross‐sectional study. The esophageal body DP and compliance were measured using a customized analysis program, with compliance calculated as (Δ esophageal body volume)/(Δ pressure) between two FLIP‐filled volumes.
Results
In controls, the median (5–95th percentile) DP was 19.8 mm (17.9–21) and esophageal body compliance was 0.37 ml/mmHg (0.18–1.1), which was greater than in EoE (DP 19 (11–21)), compliance 0.19 (0.02–0.71), p ‐values 0.2 ml/mmHg) and normal DP (>17 mm); 11 (6%) had normal compliance and reduced DP; 34 (20%) had reduced compliance and normal DP; and 56 (33%) had reduced compliance and reduce DP. Patients with both reduced compliance and DP had the greatest proportion of severe rings (61% with EREFS score 2–3) and stricture (100%).
Conclusion
FLIP provides an objective evaluation of biomechanical properties of the esophageal wall that appears enhanced by complementary application of metrics of DP and esophageal body compliance.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Autor:innen und Institutionen
- Division of Gastroenterology, University of British Columbia, Vancouver General Hospital, Vancouver, British Columbia, Canada.
- Department of Medicine, Division of Gastroenterology and Hepatology, Feinberg School of Medicine, Northwestern University, Chicago, Illinois, USA.
ORCID-Profile: Wenjun Kou, Dustin A Carlson
Förderung
- NIDDK NIH HHS
- U.S. Public Health Service
Zitieren
APA
Moosavi, S., Shehata, C., Kou, W., Hirano, I., Gonsalves, N., Peterson, S., Pandolfino, J. E., & Carlson, D. A. (2023). Measuring esophageal compliance using functional lumen imaging probe to assess remodeling in eosinophilic esophagitis. Neurogastroenterology & Motility, 35(4). https://doi.org/10.1111/nmo.14525
Vancouver
Moosavi S, Shehata C, Kou W, Hirano I, Gonsalves N, Peterson S, et al. Measuring esophageal compliance using functional lumen imaging probe to assess remodeling in eosinophilic esophagitis. Neurogastroenterology & Motility. 2023;35(4). doi:10.1111/nmo.14525
BibTeX
@article{Moosavi2023,
author = {Moosavi, Sarvee and Shehata, Christina and Kou, Wenjun and Hirano, Ikuo and Gonsalves, Nirmala and Peterson, Stephanie and Pandolfino, John E and Carlson, Dustin A},
title = {Measuring esophageal compliance using functional lumen imaging probe to assess remodeling in eosinophilic esophagitis},
journal = {Neurogastroenterology & Motility},
year = {2023},
volume = {35},
number = {4},
doi = {10.1111/nmo.14525}
}
EUVIMED