Identification of Filamentous Fungi: An Evaluation of Three MALDI ‐ TOF Mass Spectrometry Systems
Mycoses · 2025-06 · 68 (6)
DOI: 10.1111/myc.70078 ↗Kurz gefasst
Evaluationsstudie in der Zeitschrift Mycoses, online erschienen am 23.06.2025. Verfasst von Isabel Klugherz und 6 weiteren Autor:innen. Beteiligte Einrichtung: Diagnostic and Research Institute of Hygiene, Microbiology and Environmental Medicine, Medical University of Graz, Graz, Austria. Hauptthemen laut MEDLINE: Spectrometry, Mass, Matrix-Assisted Laser Desorption-Ionization, Fungi. Die Arbeit wurde bisher 9-mal zitiert (Crossref) und nennt 29 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).
Abstract (Originalsprache)
Background
Identification of filamentous fungi still poses a major challenge to laboratories. Matrix assisted laser desorption/ionisation time of flight mass spectrometry (MALDI‐TOF MS) is a promising tool, since it offers low‐cost and fast results. Different MALDI‐TOF MS systems are available for routine laboratories. This is a comprehensive head‐to‐head comparison of three devices and the respective reference spectrum databases. Materials and Methods A set of 77 pre‐characterised isolates of filamentous fungi was measured (in duplicates) parallelly with three MALDI‐TOF MS systems after 24 h, 48 h and 72 h of incubation: (1) Biotyper smart (‘BT’, Bruker Daltonics),(2) EXS2600 (‘EXS’, Zybio) and (3) VITEK MS PRIME (‘VITEK’, bioMérieux).
Results
After three measurements, no valid results at the species level (‘green category’) were obtained for 18%, 21% and 14% of isolates by BT, EXS and VITEK. Depending on the MALDI‐TOF MS system, validity rates ranged from 58%–82% for the different time points. Correct and valid results were obtained for 82%, 73% and 81% of isolates by BT, EXS and VITEK. BT was the system that required the most duplicate measurements. EXS displayed the highest rate of misidentification events. VITEK had the highest rate of unidentified isolates, which were featured within its database.
Conclusion
Based on our experience, all three devices proved to be suitable for routine diagnostics. The timepoint of measurement had a major impact on the quality of analysis and should be considered by the user. Lowering the validity cut‐off levels might increase the performance of the EXS system.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Diagnostic and Research Institute of Hygiene, Microbiology and Environmental Medicine, Medical University of Graz, Graz, Austria.
- Division of Clinical Microbiology, Department of Laboratory Medicine, Medical University of Vienna, Vienna, Austria.
ORCID-Profile: Birgit Willinger, Karl Dichtl
Zitieren
APA
Klugherz, I., Kölli, B., Reinmüller, M., Willinger, B., Leitner, E., Steinmetz, I., & Dichtl, K. (2025). Identification of Filamentous Fungi: An Evaluation of Three MALDI ‐ TOF Mass Spectrometry Systems. Mycoses, 68(6). https://doi.org/10.1111/myc.70078
Vancouver
Klugherz I, Kölli B, Reinmüller M, Willinger B, Leitner E, Steinmetz I, et al. Identification of Filamentous Fungi: An Evaluation of Three MALDI ‐ TOF Mass Spectrometry Systems. Mycoses. 2025;68(6). doi:10.1111/myc.70078
BibTeX
@article{Klugherz2025,
author = {Klugherz, Isabel and Kölli, Bettina and Reinmüller, Moritz and Willinger, Birgit and Leitner, Eva and Steinmetz, Ivo and Dichtl, Karl},
title = {Identification of Filamentous Fungi: An Evaluation of Three MALDI ‐ TOF Mass Spectrometry Systems},
journal = {Mycoses},
year = {2025},
volume = {68},
number = {6},
doi = {10.1111/myc.70078}
}
EUVIMED