EUVIMEDEUVIMED entdecken
EvaluationsstudieOpen Access · CC BY 4.0

Droplet Digital PCR: A Powerful Tool for Accurate Quantification of Hepatitis D Virus RNA Levels and Verification of Detection Limits

L Sandmann iD, L Windzio, B Bremer, S Falak, J Beheim-Schwarzbach, A Kummrow, M Cornberg iD, H Wedemeyer iD, B Maasoumy, E Valiente

Journal of Viral Hepatitis · 2025

DOI: 10.1111/jvh.70036 ↗

Kurz gefasst

Evaluationsstudie in der Zeitschrift Journal of Viral Hepatitis, 2025. Verfasst von L Sandmann und 9 weiteren Autor:innen. Beteiligte Einrichtung: Department of Gastroenterology, Hepatology, Infectious Diseases and Endocrinology, Hannover Medical School, Hannover, Germany. Hauptthemen laut MEDLINE: Hepatitis Delta Virus, RNA, Viral, Hepatitis D, Viral Load. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 31 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: European Association of National Metrology Institutes.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Reliable quantification of hepatitis D virus (HDV) RNA levels is necessary for initiating and guiding antiviral treatment. The aim of this work is to develop and validate a digital PCR method for the accurate quantification of HDV RNA, including evaluation of its clinical accuracy, especially for low‐concentrated clinical samples. The reverse transcription digital PCR (RT‐dPCR) development followed the standard procedure, including primer design, determination of linearity, calculation of recovery and the intermediate precision of the RNA extraction kits, determination of the limit of detection (LOD) and quantification (LOQ), droplet size measurements, conversion factor, and uncertainty budget. The World Health Organisation (WHO)‐HDV international standard was used for RT‐dPCR development. Commutability of the new method was explored, comparing RT‐dPCR with quantification assays applied in clinical routine using clinical plasma samples covering a range of HDV RNA concentrations. The conversion factor from copies/mL to IU/mL was 0.77. LOD and LOQ of the RT‐dPCR were 0.7 copies/mL (0.56 IU/mL) and 10 copies/mL (8 IU/mL), respectively. When evaluating the qualitative results of the clinical HDV samples at low concentrations, 31% of the HDV clinical samples tested negative by RT‐qPCR were tested positive by RT‐dPCR. The RT‐qPCR and RT‐dPCR quantitative data showed a good correlation with a standard deviation of ±1.12 log IU/mL. RT‐dPCR is an accurate method for HDV RNA quantification that may serve as a complement to RT‐qPCR, especially when accurate detection is essential for decision making in clinical settings.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Hepatitis Delta VirusRNA, ViralHepatitis DViral LoadReverse Transcriptase Polymerase Chain ReactionReal-Time Polymerase Chain ReactionHumansLimit of DetectionSensitivity and Specificity

Stichwörter der Autor:innen

HDVHDV RNART‐dPCRRT‐qPCRquantificationverification

Substanzen

RNA, Viral

Autor:innen und Institutionen

  • Department of Gastroenterology, Hepatology, Infectious Diseases and Endocrinology, Hannover Medical School, Hannover, Germany.
  • Excellence Cluster RESIST, Excellence Initiative Hannover Medical School, Hannover, Germany.
  • D-SOLVE Consortium, an EU Horizon Europe Funded Project (No. 101057917), Hannover, Germany.
  • Department 8.3, Biomedical Optics, Physikalisch-Technische Bundesanstalt, Berlin, Germany.
  • German Center for Infection Research (DZIF), Hannover-Braunschweig, Germany.
  • Center for Individualised Infection Medicine (CiiM), Hannover Medical School/Helmholtz Center, Hannover, Germany.

ORCID-Profile: L Sandmann, M Cornberg, H Wedemeyer

Förderung

  • European Association of National Metrology Institutes

Zitieren

APA

Sandmann, L., Windzio, L., Bremer, B., Falak, S., Beheim-Schwarzbach, J., Kummrow, A., Cornberg, M., Wedemeyer, H., Maasoumy, B., & Valiente, E. (2025). Droplet Digital PCR: A Powerful Tool for Accurate Quantification of Hepatitis D Virus RNA Levels and Verification of Detection Limits. Journal of Viral Hepatitis. https://doi.org/10.1111/jvh.70036

Vancouver

Sandmann L, Windzio L, Bremer B, Falak S, Beheim-Schwarzbach J, Kummrow A, et al. Droplet Digital PCR: A Powerful Tool for Accurate Quantification of Hepatitis D Virus RNA Levels and Verification of Detection Limits. Journal of Viral Hepatitis. 2025. doi:10.1111/jvh.70036

BibTeX
@article{Sandmann2025,
  author = {Sandmann, L and Windzio, L and Bremer, B and Falak, S and Beheim-Schwarzbach, J and Kummrow, A and Cornberg, M and Wedemeyer, H and Maasoumy, B and Valiente, E},
  title = {Droplet Digital PCR: A Powerful Tool for Accurate Quantification of Hepatitis D Virus RNA Levels and Verification of Detection Limits},
  journal = {Journal of Viral Hepatitis},
  year = {2025},
  doi = {10.1111/jvh.70036}
}
Kostenlos starten →