EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Accuracy and agreement of a wearable device for heart rate and oxygen saturation monitoring in awake dogs

Ş Melek, Ö T Orhun, M G Şenocak, T Arslan

Journal of Small Animal Practice · 2026-11 · 67 (11) · 1106-1111

DOI: 10.1111/jsap.70162 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Small Animal Practice, online erschienen am 25.06.2026. Verfasst von Ş Melek und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Surgery, Faculty of Veterinary Medicine, Bingol University, Bingöl, Turkey. Hauptthemen laut MEDLINE: Heart Rate, Oxygen Saturation, Wearable Electronic Devices, Oximetry. Die Arbeit nennt 19 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Türkiye Bilimsel ve Teknolojik Araştırma Kurumu.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objectives

The present study evaluated the accuracy and agreement of heart rate and peripheral oxygen saturation measurements obtained using the Owlet Smart Sock in awake dogs with low activity levels, using a veterinary patient monitor as the reference method. Materials and Methods Clinically healthy dogs underwent simultaneous heart rate and peripheral oxygen saturation measurements with Owlet Smart Sock and a veterinary patient monitor during routine preoperative evaluation. Method comparison was performed using Pearson correlation analysis and Bland–Altman agreement analysis. Clinical equivalence was assessed using the two one‐sided tests procedure with predefined equivalence margins of ±10 bpm for heart rate and ±3% for peripheral oxygen saturation. A total of 237 paired measurements were analysed.

Results

Heart rate measurements obtained with Owlet Smart Sock showed a strong positive linear relationship between the two methods ( r = 0.934, P < .001). Bland–Altman analysis demonstrated a mean bias of −0.46 bpm, with narrow limits of agreement. Two one‐sided tests analysis indicated that the mean difference in heart rate measurements lay within the predefined equivalence margin ( P < .001). In contrast, peripheral oxygen saturation measurements showed no significant correlation between devices ( r = 0.038, P = .562), and Bland–Altman analysis revealed considerable variability at the individual measurement level. Clinical Significance The Owlet Smart Sock demonstrated acceptable agreement with the reference monitor for heart rate measurement in awake dogs with low activity levels, whereas peripheral oxygen saturation measurements showed limited agreement and should be interpreted with caution in clinical practice.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Heart RateOxygen SaturationWearable Electronic DevicesOximetryAnimalsDogsFemaleMaleMonitoring, PhysiologicReproducibility of Results

Autor:innen und Institutionen

  • Department of Surgery, Faculty of Veterinary Medicine, Bingol University, Bingöl, Turkey.
  • Department of Surgery, Faculty of Eregli Veterinary Medicine, Necmettin Erbakan University, Konya, Turkey.
  • Department of Surgery, Faculty of Veterinary Medicine, Atatürk University, Erzurum, Turkey.

Förderung

  • Türkiye Bilimsel ve Teknolojik Araştırma Kurumu

Zitieren

APA

Melek, Ş., Orhun, Ö. T., Şenocak, M. G., & Arslan, T. (2026). Accuracy and agreement of a wearable device for heart rate and oxygen saturation monitoring in awake dogs. Journal of Small Animal Practice, 67(11), 1106-1111. https://doi.org/10.1111/jsap.70162

Vancouver

Melek Ş, Orhun ÖT, Şenocak MG, Arslan T. Accuracy and agreement of a wearable device for heart rate and oxygen saturation monitoring in awake dogs. Journal of Small Animal Practice. 2026;67(11):1106-1111. doi:10.1111/jsap.70162

BibTeX
@article{Melek2026,
  author = {Melek, Ş and Orhun, Ö T and Şenocak, M G and Arslan, T},
  title = {Accuracy and agreement of a wearable device for heart rate and oxygen saturation monitoring in awake dogs},
  journal = {Journal of Small Animal Practice},
  year = {2026},
  volume = {67},
  number = {11},
  pages = {1106--1111},
  doi = {10.1111/jsap.70162}
}
Kostenlos starten →