Initial retention force of three attachment systems for implant retained‐mandibular overdentures: An in vitro study
Journal of Prosthodontics · 2024-06 · 33 (5) · 475-483
DOI: 10.1111/jopr.13729 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift Journal of Prosthodontics, 2023-08. Verfasst von Raphael Wakam und 3 weiteren Autor:innen. Beteiligte Einrichtung: Unité de Recherche Biomatériaux Innovants et Interfaces, Faculty of odontology, Université Paris Cité, Montrouge, France. Hauptthemen laut MEDLINE: Denture, Overlay, Denture Retention, Dental Prosthesis, Implant-Supported, Dental Stress Analysis. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 31 Literaturangaben.
Abstract (Originalsprache)
Purpose
To evaluate and compare the initial retention force of three resilient unsplinted attachment systems for implant‐retained mandibular overdentures: two cylindrical attachment systems (Locator R‐Tx® and Novaloc®), and one ball attachment system (Ball System). Materials and Methods For each attachment system, initial retention is measured as the average of the maximal dislodging forces during 10 insertion‐removal cycles. For the Ball System, three activation degrees of the matrix are included versus four and six color‐coded retention devices for the Locator R‐Tx® and the Novaloc®, respectively, to represent the complete regular retention devices panel. For each retention device or activation degree, eight samples are tested.
Results
The initial retention range is similar between the Ball System (7.7 ± 3.4 N – 19.9 ± 4.6 N) and the Novaloc® (2.0 ± 0.5 N – 18.9 ± 1.4 N) and broader for the Locator R‐Tx® (3.3 ± 5.0 N – 60.2 ± 6.0 N). In each attachment system, the initial retention of each retention device is significantly different from the others, except for the two most retentive Novaloc® ones. Retention devices were also classified according to their initial retention (low, medium, and maximum). In each retention group, the Novaloc® and the Ball System provided similar retention values lower than the Locator R‐Tx®.
Conclusion
Most of the retention devices tested provided an initial retention force of over 5 N for all three attachment systems. The Locator R‐Tx® had the most comprehensive range, and the Novaloc® seemed to provide the most reproducible values, unlike the Ball System due to the activation required by the operator.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Unité de Recherche Biomatériaux Innovants et Interfaces, Faculty of odontology, Université Paris Cité, Montrouge, France.
- Prosthetics Department, Oral Medicine Service, Pôle des Spécialités médico-chirurgicales, CHU Lille, France.
- Université Sorbonne Paris Nord, Bobigny, France.
- Laboratory URP2496 Orofacial Pathologies, Imaging and Biotherapies, Faculty of odontology, Université Paris Cité, Montrouge, France.
- Oral Medicine Service, Prosthetics Department, AP-HP/GH Nord, Paris, France.
ORCID-Profile: Caroline Gorin
Zitieren
APA
Wakam, R., Mawussi, K. B., Benoit, A., & Gorin, C. (2024). Initial retention force of three attachment systems for implant retained‐mandibular overdentures: An in vitro study. Journal of Prosthodontics, 33(5), 475-483. https://doi.org/10.1111/jopr.13729
Vancouver
Wakam R, Mawussi KB, Benoit A, Gorin C. Initial retention force of three attachment systems for implant retained‐mandibular overdentures: An in vitro study. Journal of Prosthodontics. 2024;33(5):475-483. doi:10.1111/jopr.13729
BibTeX
@article{Wakam2024,
author = {Wakam, Raphael and Mawussi, Kwamivi B and Benoit, Aurélie and Gorin, Caroline},
title = {Initial retention force of three attachment systems for implant retained‐mandibular overdentures: An in vitro study},
journal = {Journal of Prosthodontics},
year = {2024},
volume = {33},
number = {5},
pages = {475--483},
doi = {10.1111/jopr.13729}
}
EUVIMED