Portable Ultra‐Low‐Field MRI in Outpatient Neurology: An Examination of Clinical Performance and Patient Experience
Journal of Neuroimaging · 2026
DOI: 10.1111/jon.70143 ↗Kurz gefasst
Multizentrische Studie in der Zeitschrift Journal of Neuroimaging, 2026. Verfasst von Laszlo Mechtler und 6 weiteren Autor:innen. Beteiligte Einrichtung: Dent Neurologic Institute, Buffalo, New York, USA. Hauptthemen laut MEDLINE: Ambulatory Care, Brain Diseases, Magnetic Resonance Imaging, Patient Satisfaction. Die Arbeit nennt 41 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).
Abstract (Originalsprache)
Background and Purpose Brain magnetic resonance imaging (MRI) is an essential component for outpatient neurological evaluation, though access to timely imaging at the point of care is not feasible for most outpatient neurology practices. Portable ultra‐low‐field MRI systems offer a potential solution, but studies describing their clinical performance in routine outpatient neurology practice remain limited. This study evaluated clinical concordance and patient experience of portable MRI (pMRI) compared to standard‐of‐care MRI (sMRI) in independent neurology practices.
Methods
In this prospective, multicenter study, adults presenting to outpatient neurology clinics and requiring clinically indicated brain MRI completed imaging with both pMRI (0.064 T) and sMRI (91% 3 T, 9% 1.5 T) in the neurology clinic. All examinations were independently reviewed by board‐certified neuroradiologists in a blinded, randomized fashion without access to clinical history. The primary endpoint was patient‐level concordance between modalities for the presence or absence of abnormal findings. Discordant cases underwent post hoc unblinded paired review with clinical history to assess clinical significance. Secondary endpoints included patient‐reported experience, assessed using a structured questionnaire evaluating noise, comfort, claustrophobia, anxiety, and overall experience on 10‐point Likert scales, as well as modality preference.
Results
Among 125 participants imaged for common outpatient indications, including headache (40%), cognitive impairment or dementia (16%), multiple sclerosis follow‐up (16%), and tumor surveillance (16%), portable and sMRI demonstrated 92% concordance on blinded review. Following clinically informed post hoc review, concordance increased to 98%. pMRI was rated more favorably across all patient experience domains ( p < 0.001), and participants preferred pMRI (61%) over sMRI (14%) by a 4:1 ratio.
Conclusions
In outpatient neurology practices, pMRI images demonstrated high clinical concordance with sMRI for identifying the presence or absence of structural brain abnormalities and were strongly preferred by patients. These findings support the use of pMRI as a practical point‐of‐care imaging tool for neurology patients, enabling timely access to structural neuroimaging during the clinical encounter.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Dent Neurologic Institute, Buffalo, New York, USA.
- Departments of Neurology, Oncology, & Neurosurgery at the Jacobs School of Medicine and Biomedical Sciences, State University of New York, Buffalo, New York, USA.
- Department of Radiology (Interventional Neuroradiology), Neurological Surgery, Ken and Ruth Davee Department of Neurology, Northwestern University Feinberg School of Medicine, Chicago, Illinois, USA.
- Department of Clinical Radiology, University of Pennsylvania, Philadelphia, Pennsylvania, USA.
- Hyperfine Inc., Guilford, Connecticut, USA.
- Department of Radiology, Jacobs School of Medicine and Biomedical Sciences, State University of New York, Buffalo, New York, USA.
Zitieren
APA
Mechtler, L., Ansari, S. A., Derakhshan, J., Knopp, E., Pinter, N., Nasri, M. A., & Samuel, G. (2026). Portable Ultra‐Low‐Field MRI in Outpatient Neurology: An Examination of Clinical Performance and Patient Experience. Journal of Neuroimaging. https://doi.org/10.1111/jon.70143
Vancouver
Mechtler L, Ansari SA, Derakhshan J, Knopp E, Pinter N, Nasri MA, et al. Portable Ultra‐Low‐Field MRI in Outpatient Neurology: An Examination of Clinical Performance and Patient Experience. Journal of Neuroimaging. 2026. doi:10.1111/jon.70143
BibTeX
@article{Mechtler2026,
author = {Mechtler, Laszlo and Ansari, Sameer Ahmad and Derakhshan, Jamal and Knopp, Edmond and Pinter, Nandor and Nasri, Mohamad Asaad and Samuel, Gincy},
title = {Portable Ultra‐Low‐Field MRI in Outpatient Neurology: An Examination of Clinical Performance and Patient Experience},
journal = {Journal of Neuroimaging},
year = {2026},
doi = {10.1111/jon.70143}
}