Semiautomatic Quantification of 99m Tc‐TRODAT‐1 SPECT Images in Patients With Idiopathic Parkinson's Disease
Journal of Neuroimaging · 2025-03 · 35 (2)
DOI: 10.1111/jon.70038 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Neuroimaging, online erschienen am 25.03.2025. Verfasst von Gary Ka Wai Chan und 5 weiteren Autor:innen. Beteiligte Einrichtung: Nuclear Medicine Unit, Department of Radiology, Queen Mary Hospital, Hong Kong, SAR, China. Hauptthemen laut MEDLINE: Parkinson Disease, Tomography, Emission-Computed, Single-Photon, Tropanes, Organotechnetium Compounds. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 27 Literaturangaben.
Abstract (Originalsprache)
Background and Purpose 99m Tc‐TRODAT‐1 SPECT imaging is an imaging technique, more commonly used in Asia, to diagnose Parkinson's disease (PD). This study evaluates the use of automated three‐dimensional volume‐of‐interest (VOI) analysis in diagnosing PD.
Methods
99m Tc‐TRODAT‐1 SPECT images of 76 patients (50 with PD and 26 without PD) were retrospectively analyzed. The specific binding ratio (SBR) was calculated using an automated program. Multiple linear regression and receiver operating characteristic curve analyses were performed to identify the factors that affect SBR value and determine the optimal cutoff values.
Results
Multiple regression analysis revealed disease status as the strongest predictor of SBR values, followed by age and sex. Receiver operating characteristic curve analysis demonstrated good diagnostic performance for the striatum (area under the curve [AUC] = 0.922), putamen (AUC = 0.922), and caudate (AUC = 0.881). Optimal cutoff values were determined for the striatum (0.515; sensitivity 88.5%, specificity 90.0%), putamen (0.445; sensitivity 92.3%, specificity 86.0%), and caudate (0.655; sensitivity 84.6%, specificity 90.0%).
Conclusions
Semiautomatic quantitative analysis of 99m Tc‐TRODAT‐1 SPECT using automated three‐dimensional VOI shows excellent diagnostic performance in differentiating PD from non‐Parkinson's cases. Age and sex significantly influence SBR values, suggesting the need for demographic‐adjusted cutoff values in clinical practice.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Nuclear Medicine Unit, Department of Radiology, Queen Mary Hospital, Hong Kong, SAR, China.
- Nuclear Medicine Unit, Department of Radiology and Nuclear Medicine, Tuen Mun Hospital, Hong Kong, SAR, China.
- Division of Neurology, Department of Medicine, Queen Mary Hospital, University of Hong Kong, Hong Kong, SAR, China.
ORCID-Profile: Gary Ka Wai Chan, Tsz Kit Chow, William C Y Leung
Zitieren
APA
Chan, G. K. W., Chow, T. K., Ho, R. W. H., Leung, W. C. Y., Hui, Y. H., & Ho, W. Y. (2025). Semiautomatic Quantification of 99m Tc‐TRODAT‐1 SPECT Images in Patients With Idiopathic Parkinson's Disease. Journal of Neuroimaging, 35(2). https://doi.org/10.1111/jon.70038
Vancouver
Chan GKW, Chow TK, Ho RWH, Leung WCY, Hui YH, Ho WY. Semiautomatic Quantification of 99m Tc‐TRODAT‐1 SPECT Images in Patients With Idiopathic Parkinson's Disease. Journal of Neuroimaging. 2025;35(2). doi:10.1111/jon.70038
BibTeX
@article{Chan2025,
author = {Chan, Gary Ka Wai and Chow, Tsz Kit and Ho, Ryan Wui Hang and Leung, William C Y and Hui, Yan Ho and Ho, Wai Yin},
title = {Semiautomatic Quantification of 99m Tc‐TRODAT‐1 SPECT Images in Patients With Idiopathic Parkinson's Disease},
journal = {Journal of Neuroimaging},
year = {2025},
volume = {35},
number = {2},
doi = {10.1111/jon.70038}
}
EUVIMED