A novel fully automated method for measuring ASPECTS to improve stroke diagnosis: Comparison to traditional ASPECTS
Journal of Neuroimaging · 2024
DOI: 10.1111/jon.13159 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Neuroimaging, 2024. Verfasst von Georg Gohla und 9 weiteren Autor:innen. Beteiligte Einrichtung: Department of Diagnostic and Interventional Neuroradiology, Eberhard Karls-University Tübingen, Tübingen, Germany. Hauptthemen laut MEDLINE: Stroke, Brain Ischemia. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 26 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0). Gefördert von: None., None.
Themen: Schlaganfall
Abstract (Originalsprache)
Background and Purpose To compare the accuracy of subjective Alberta Stroke Program Early CT Score (sASPECTS) evaluation and that of an automated prototype software (aASPECTS) on nonenhanced CT (NECT) in patients with early anterior territory stroke and controls using side‐to‐side quantification of hypoattenuated brain areas.
Methods
We retrospectively analyzed the NECT scans of 42 consecutive patients with ischemic stroke before reperfusion and 42 controls using first sASPECTS and subsequently aASPECTS. We assessed the differences in Alberta Stroke Program Early CT Score (ASPECTS) and calculated the sensitivity and specificity of NECT with CT perfusion, whereas cerebral blood volume (CBV) served as the reference standard for brain infarction.
Results
The clot was located in the middle cerebral artery (MCA) in 47.6% of cases and the internal carotid artery (ICA) in 28.6% of cases. Ten cases presented combined ICA and MCA occlusions. The stroke was right sided in 52.4% of cases and left sided in 47.6%. Reader‐based NECT analysis yielded a median sASPECTS of 10. The median CBV‐based ASPECTS was 7. Compared to the area of decreased CBV, sASPECTS yielded a sensitivity of 12.5% and specificity of 86.8%. The software prototype (aASPECTS) yielded an overall sensitivity of 65.5% and a specificity of 92.2%. The interreader agreement for ASPECTS evaluation of admission NECT and follow‐up CT was almost perfect ( κ = .93). The interreader agreement of the CBV color map evaluation was substantial ( κ = .77). Conclusions aASPECTS of NECT can outperform sASPECTS for stroke detection.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Diagnostic and Interventional Neuroradiology, Eberhard Karls-University Tübingen, Tübingen, Germany.
- Department of Diagnostic and Interventional Radiology, Eberhard Karls-University Tübingen, Tübingen, Germany.
- Radiology Salzstraße, Muenster, Germany.
- Siemens AG Healthcare; Imaging & Therapy Systems Computed Tomography & Radiation Oncology, Forchheim, Germany.
- Russell H. Morgan Department of Radiology and Radiological Science, Johns Hopkins University School of Medicine, Baltimore, Maryland, USA.
- Institute of Neuroradiology, Philipps University of Marburg, Marburg, Germany.
ORCID-Profile: Georg Gohla, Arne Estler
Förderung
- None.
- None
Zitieren
APA
Gohla, G., Schwarz, R., Bier, G., Estler, A., Bongers, M. N., Ditt, H., Fritz, J., Kemmling, A., Ernemann, U., & Horger, M. (2024). A novel fully automated method for measuring ASPECTS to improve stroke diagnosis: Comparison to traditional ASPECTS. Journal of Neuroimaging. https://doi.org/10.1111/jon.13159
Vancouver
Gohla G, Schwarz R, Bier G, Estler A, Bongers MN, Ditt H, et al. A novel fully automated method for measuring ASPECTS to improve stroke diagnosis: Comparison to traditional ASPECTS. Journal of Neuroimaging. 2024. doi:10.1111/jon.13159
BibTeX
@article{Gohla2024,
author = {Gohla, Georg and Schwarz, Ricarda and Bier, Georg and Estler, Arne and Bongers, Malte N and Ditt, Hendrik and Fritz, Jan and Kemmling, André and Ernemann, Ulrike and Horger, Marius},
title = {A novel fully automated method for measuring ASPECTS to improve stroke diagnosis: Comparison to traditional ASPECTS},
journal = {Journal of Neuroimaging},
year = {2024},
doi = {10.1111/jon.13159}
}