Comparison of Laser Doppler Flowmetry With Contrast‐Enhanced Ultrasound to Approximate Placental Microvascular Blood Flow in the African Green Monkey ( Chlorocebus aethiops sabaeus )
Journal of Medical Primatology · 2026
DOI: 10.1111/jmp.70059 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift Journal of Medical Primatology, 2026. Verfasst von Yarines Gonzalez-Rodriguez und 6 weiteren Autor:innen. Beteiligte Einrichtung: Animal Resources Program, Wake Forest University School of Medicine, Winston-Salem, North Carolina, USA. Hauptthemen laut MEDLINE: Placenta, Laser-Doppler Flowmetry, Placental Circulation. Die Arbeit nennt 63 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NHLBI NIH HHS, NIH Office of the Director, U.S. Department of Defense.
Abstract (Originalsprache)
Background
The ability to study in vivo microvascular flow within the placenta is limited. We sought to compare two techniques for evaluation of placental perfusion in a translational nonhuman primate model.
Methods
We measured placental microvascular perfusion in six pregnant African green monkeys ( Chlorocebus aethiops sabaeus ) using both Laser Doppler Flowmetry (LDF) and contrast‐enhanced ultrasound (CEUS) in 2–4 discrete placental locations per animal at early‐ (EG), mid‐ (MG), and/or late‐gestation (LG). We assessed correlation between matched LDF and CEUS, with the hypothesis that we would observe concordance between these two measures of relative perfusion.
Results
Using 35 total paired measurements (13 EG, n = 4; 12 MG, n = 4; 10 LG, n = 5), we found no correlation between LDF microvascular perfusion and CEUS microvascular flux rate (Pearson r = −0.2026, R 2 = 0.0411, p = 0.2431).
Conclusion
LDF and CEUS produce discordant results with respect to placental perfusion. LDF is ill‐suited to a heterogeneous organ such as the placenta.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Animal Resources Program, Wake Forest University School of Medicine, Winston-Salem, North Carolina, USA.
- Lampe Joint Department of Biomedical Engineering, University of North Carolina-North Carolina State University (UNC-NC State), Raleigh, North Carolina, USA.
- Department of Pathology, Wake Forest University School of Medicine, Winston-Salem, North Carolina, USA.
- Department of Pediatrics, Wake Forest University School of Medicine, Winston-Salem, North Carolina, USA.
ORCID-Profile: Sarah N Cilvik
Förderung
- NHLBI NIH HHS
- NIH Office of the Director
- U.S. Department of Defense
- NIGMS NIH HHS
- National Heart, Lung, and Blood Institute
- National Institute of General Medical Sciences
Zitieren
APA
Gonzalez-Rodriguez, Y., Walmer, R. W., Krainiak, S., Carter, K., Kavanagh, K., Johnson, K., & Cilvik, S. N. (2026). Comparison of Laser Doppler Flowmetry With Contrast‐Enhanced Ultrasound to Approximate Placental Microvascular Blood Flow in the African Green Monkey ( Chlorocebus aethiops sabaeus ). Journal of Medical Primatology. https://doi.org/10.1111/jmp.70059
Vancouver
Gonzalez-Rodriguez Y, Walmer RW, Krainiak S, Carter K, Kavanagh K, Johnson K, et al. Comparison of Laser Doppler Flowmetry With Contrast‐Enhanced Ultrasound to Approximate Placental Microvascular Blood Flow in the African Green Monkey ( Chlorocebus aethiops sabaeus ). Journal of Medical Primatology. 2026. doi:10.1111/jmp.70059
BibTeX
@article{GonzalezRodriguez2026,
author = {Gonzalez-Rodriguez, Yarines and Walmer, Rachel W and Krainiak, Shannon and Carter, Kelli and Kavanagh, Kylie and Johnson, Kennita and Cilvik, Sarah N},
title = {Comparison of Laser Doppler Flowmetry With Contrast‐Enhanced Ultrasound to Approximate Placental Microvascular Blood Flow in the African Green Monkey ( Chlorocebus aethiops sabaeus )},
journal = {Journal of Medical Primatology},
year = {2026},
doi = {10.1111/jmp.70059}
}
EUVIMED