EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Reduced order modelling of air‐puff test for corneal material characterisation

Osama M Maklad iD, Muting Hao iD

Journal of Microscopy · 10.03.2026

DOI: 10.1111/jmi.70077 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Microscopy, online erschienen am 10.03.2026. Verfasst von Osama M Maklad und 1 weiteren Autor:innen. Beteiligte Einrichtung: Centre for Advanced Manufacturing and Materials, University of Greenwich, London, UK. Die Arbeit nennt 27 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: EPSRC National Fellowships in Fluid Dynamics (NFFDy) Summer programme 2024.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Models of the fluid–structure interaction (FSI) model for the air‐puff test were analysed. Using Abaqus, the air‐puff test is applied to eyes with varying biomechanical parameters, such as material properties, corneal thickness, and radius. A reduced order model of the air puff (a turbulent impinging jet) has been acquired to decrease simulation time from 48 h for the FSI model to approximately 12 min for the finite element analysis (FEA) model alone. To further accelerate simulations and improve model accuracy, Physics‐Informed Neural Networks (PINNs) will be integrated with the reduced‐order model. This hybrid approach will help expand the model to a larger dataset, enhancing intraocular pressure (IOP) estimation accuracy and the corneal material properties algorithm through inverse FEA. Additionally, a neural network (NN) framework with embedded Gaussian‐modulated waveforms is proposed to model the pressure and deformation distributions on the corneal surface as functions of spatial and temporal parameters. By learning the relationship between corneal biomechanical inputs such as corneal central thickness (CCT), intraocular pressure (IOP), and baseline properties (), and the governing coefficients of pressure and deformation, the network accurately reconstructs the result that matches well with the high‐fidelity CFD data. This approach can quickly capture the distribution of pressure and deformation. It can also provide insights into the distinct spatial and temporal dynamics of pressure and deformation, giving a more comprehensive understanding of fluid–structure interaction phenomena in the air‐puff test.

Schlagwörter

Stichwörter der Autor:innen

Gradient Boosting Regressor (GBR)air‐puff testfluid–structure interaction (FSI)intraocular pressure (IOP)machine learning (ML)ocular biomechanicsreduced order modelling

Autor:innen und Institutionen

  • Centre for Advanced Manufacturing and Materials, University of Greenwich, London, UK.
  • Oxford Thermofluid Institute, St John's College, University of Oxford, Oxford, UK.

ORCID-Profile: Osama M Maklad, Muting Hao

Förderung

  • EPSRC National Fellowships in Fluid Dynamics (NFFDy) Summer programme 2024

Zitieren

APA

Maklad, O. M., & Hao, M. (2026). Reduced order modelling of air‐puff test for corneal material characterisation. Journal of Microscopy. https://doi.org/10.1111/jmi.70077

Vancouver

Maklad OM, Hao M. Reduced order modelling of air‐puff test for corneal material characterisation. Journal of Microscopy. 2026. doi:10.1111/jmi.70077

BibTeX
@article{Maklad2026,
  author = {Maklad, Osama M and Hao, Muting},
  title = {Reduced order modelling of air‐puff test for corneal material characterisation},
  journal = {Journal of Microscopy},
  year = {2026},
  doi = {10.1111/jmi.70077}
}
Kostenlos starten →