Contamination of Disposable Distal Cap Duodenoscopes and Detachable Elevator Duodenoscopes After Reprocessing: A Randomized Trial
Journal of Gastroenterology and Hepatology · 2025
DOI: 10.1111/jgh.16827 ↗Kurz gefasst
Randomisierte kontrollierte Studie in der Zeitschrift Journal of Gastroenterology and Hepatology, 2025. Verfasst von Wiriyaporn Ridtitid und 9 weiteren Autor:innen. Beteiligte Einrichtung: Division of Gastroenterology, Department of Medicine, Faculty of Medicine, Chulalongkorn University, Bangkok, Thailand. Hauptthemen laut MEDLINE: Duodenoscopes, Equipment Contamination, Disinfection, Disposable Equipment. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 30 Literaturangaben. Gefördert von: Thailand Science Research and Innovation Fund Chulalongkorn University, KOSIN Medical Supply Co., Ltd., Bangkok, Thailand.
Abstract (Originalsprache)
Background and Aim To reduce bacterial contamination after reprocessing, various new designs of duodenoscopes have been developed to better expose the elevator complex for cleaning. We compared the rates of bacterial contamination and organic residue in disposable distal cap duodenoscopes and detachable elevator duodenoscopes after manual cleaning and high‐level disinfection (HLD), as well as their cost‐effectiveness.
Methods
A total of 162 duodenoscopes were randomly assigned to either Group A (disposable distal caps; n = 81) or Group B (detachable elevator; n = 81). A total of 324 samples from the elevator were collected for culture following manual cleaning ( n = 81 in each group) and HLD ( n = 81 in each group), followed by the adenosine triphosphate (ATP) testing for organic residue.
Results
After manual cleaning, there was no difference in bacterial contamination rates (8.6% vs. 8.6%; p = 1.00) and mean ATP levels (164.6 ± 257.5 vs. 158.1 ± 286.1 RLUs; p = 0.88) between Groups A and B. After HLD, no bacterial contamination was observed in either group and the mean ATP levels were very low with no significant difference between the two groups (30.1 ± 45.3 vs. 37.5 ± 51.9 RLUs; p = 0.68). The expense in reprocessing (excluding the scope cost) for Group A was lower (2099 USD) than Group B (3854 USD) in providing comparable scope cleanliness.
Conclusion
After manual cleaning, the bacterial contamination rate and organic residue levels in detachable elevator duodenoscopes and disposable distal caps duodenoscopes were comparable. No bacterial contamination was detected in either type of duodenoscope after reprocessing. Apart from the initial differences in scope cost, the disposable distal cap duodenoscope had lower cost on disposable items to have comparable disinfection result.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Division of Gastroenterology, Department of Medicine, Faculty of Medicine, Chulalongkorn University, Bangkok, Thailand.
- Center of Excellence for Innovation and Endoscopy in Gastrointestinal Oncology, Faculty of Medicine, Chulalongkorn University, Bangkok, Thailand.
- Department of Microbiology and Antimicrobial Resistance and Stewardship Research Unit, Faculty of Medicine, Chulalongkorn University, Bangkok, Thailand.
- Department of Anatomy, Faculty of Medicine, Chulalongkorn University, Bangkok, Thailand.
- Excellence Center for Gastrointestinal Endoscopy, King Chulalongkorn Memorial Hospital, Bangkok, Thailand.
ORCID-Profile: Wiriyaporn Ridtitid, Thanawat Luangsukrerk, Pradermchai Kongkam, Rungsun Rerknimitr
Förderung
- Thailand Science Research and Innovation Fund Chulalongkorn University
- KOSIN Medical Supply Co., Ltd., Bangkok, Thailand
Zitieren
APA
Ridtitid, W., Buathong, J., Chatsuwan, T., Angsuwatcharakon, P., Luangsukrerk, T., Mekaroonkamol, P., Piyachaturawat, P., Kulpatcharapong, S., Kongkam, P., & Rerknimitr, R. (2025). Contamination of Disposable Distal Cap Duodenoscopes and Detachable Elevator Duodenoscopes After Reprocessing: A Randomized Trial. Journal of Gastroenterology and Hepatology. https://doi.org/10.1111/jgh.16827
Vancouver
Ridtitid W, Buathong J, Chatsuwan T, Angsuwatcharakon P, Luangsukrerk T, Mekaroonkamol P, et al. Contamination of Disposable Distal Cap Duodenoscopes and Detachable Elevator Duodenoscopes After Reprocessing: A Randomized Trial. Journal of Gastroenterology and Hepatology. 2025. doi:10.1111/jgh.16827
BibTeX
@article{Ridtitid2025,
author = {Ridtitid, Wiriyaporn and Buathong, Jirayus and Chatsuwan, Tanittha and Angsuwatcharakon, Phonthep and Luangsukrerk, Thanawat and Mekaroonkamol, Parit and Piyachaturawat, Panida and Kulpatcharapong, Santi and Kongkam, Pradermchai and Rerknimitr, Rungsun},
title = {Contamination of Disposable Distal Cap Duodenoscopes and Detachable Elevator Duodenoscopes After Reprocessing: A Randomized Trial},
journal = {Journal of Gastroenterology and Hepatology},
year = {2025},
doi = {10.1111/jgh.16827}
}
EUVIMED