Vergleichsstudie in der Zeitschrift Journal of Evaluation in Clinical Practice, 2025. Verfasst von Elif Günay Ismailoğlu und 4 weiteren Autor:innen. Beteiligte Einrichtung: Health Sciences Faculty, Nursing Department, Izmir Bakircay University, İzmir, Turkey. Hauptthemen laut MEDLINE: Obesity, Blood Pressure Determination, Forearm, Arm. Die Arbeit nennt 24 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Rationale

Noninvasive blood pressure (BP) monitoring is very important also difficult for accurate diagnosis and monitor of obese patients.

Aims

The study aimed to assess the agreement between forearm arm BP and intra‐arterial BP values in a different body and arm positions in obese patients.

Methods

The descriptive study was carried out on 60 intensive care patients with a body mass index above 30 kg/m 2 who were monitored with invasive radial BP in the general and surgical intensive care units. BP values obtained from the upper arm and forearm with different arm and body positions were compared with intra‐arterial BP results. Bland‐Altman analysis and correlation coefficient were used for the accuracy of upper and forearm noninvasive BP reading in different positions.

Results

The best agreement was found between the forearm systolic BP in the supine and half‐sitting position with the arm down and intra‐arterial BP values. Also, the best agreement in diastolic BP was found between the half‐sitting and half‐sitting position with the arm down and intra‐arterial BP.

Conclusions

Forearm systolic BP measurement, especially in the supine and half‐sitting position with the arm below the heart level position, was best agreement with intra‐arterial measurement, regarded as the gold standard. For this reason, it is more appropriate to use forearm BP measurement in obese patients monitored in intensive care to obtain accurate results.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

ObesityBlood Pressure DeterminationForearmArmHumansMaleFemaleMiddle AgedAdultAgedBlood PressureBody Mass IndexPatient PositioningIntensive Care Units

Stichwörter der Autor:innen

blood pressureforearm blood pressureintensive careobese patientsupper arm blood pressure

Autor:innen und Institutionen

  • Health Sciences Faculty, Nursing Department, Izmir Bakircay University, İzmir, Turkey.
  • Medical Faculty, Department of Anesthesiology, Izmir Bakircay University, İzmir, Turkey.
  • Department of Anesthesiology, İzmir Bakircay Universitesi Cigli Training and Research Hospital, İzmir, Turkey.

ORCID-Profile: Elif Günay Ismailoğlu, Seda Şahan

Zitieren

APA

Günay Ismailoğlu, E., Aygün, H., Şahan, S., Ergin, E., & Geçtan, E. (2025). Comparison of forearm and intra‐arterial blood pressure measurements according to body and arm positions in obese patients. Journal of Evaluation in Clinical Practice. https://doi.org/10.1111/jep.14093

Vancouver

Günay Ismailoğlu E, Aygün H, Şahan S, Ergin E, Geçtan E. Comparison of forearm and intra‐arterial blood pressure measurements according to body and arm positions in obese patients. Journal of Evaluation in Clinical Practice. 2025. doi:10.1111/jep.14093

BibTeX
@article{GnayIsmailolu2025,
  author = {Günay Ismailoğlu, Elif and Aygün, Hakan and Şahan, Seda and Ergin, Eda and Geçtan, Eliz},
  title = {Comparison of forearm and intra‐arterial blood pressure measurements according to body and arm positions in obese patients},
  journal = {Journal of Evaluation in Clinical Practice},
  year = {2025},
  doi = {10.1111/jep.14093}
}
Kostenlos starten →