Improved positional accuracy of dental implant placement using a haptic and machine‐vision‐controlled collaborative surgery robot: A pilot randomized controlled trial
Journal of Clinical Periodontology · 2024-01 · 51 (1) · 24-32
DOI: 10.1111/jcpe.13893 ↗Kurz gefasst
Randomisierte kontrollierte Studie in der Zeitschrift Journal of Clinical Periodontology, online erschienen am 23.10.2023. Verfasst von Jun-Yu Shi und 6 weiteren Autor:innen. Beteiligte Einrichtung: Shanghai PerioImplant Innovation Center, Department of Oral and Maxillofacial Implantology, Shanghai Ninth People's Hospital, Shanghai Jiao Tong Univ…. Hauptthemen laut MEDLINE: Dental Implants, Robotics, Surgery, Computer-Assisted. Die Arbeit wurde bisher 67-mal zitiert (Crossref) und nennt 23 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0). Gefördert von: Clinical Research Program of Ninth People's Hospital affiliated Shanghai Jiao Tong University School of Medicine, National Clinical Research Center for Oral Diseases, Shanghai Innovative Research Team Award of High-Level University.
Abstract (Originalsprache)
Aim
To compare the implant accuracy, safety and morbidity between robot‐assisted and freehand dental implant placement. Materials and Methods Subjects requiring single‐site dental implant placement were recruited. Patients were randomly allocated to freehand implant placement and robot‐assisted implant placement. Differences in positional accuracy of the implant, surgical morbidity and complications were assessed. The significance of intergroup differences was tested with an intention‐to‐treat analysis and a per‐protocol (PP) analysis (excluding one patient due to calibration error).
Results
Twenty patients (with a median age of 37, 13 female) were included. One subject assigned to the robotic arm was excluded from the PP analysis because of a large calibration error due to the dislodgement of the index. For robot‐assisted and freehand implant placement, with the PP analysis, the median (25th–75th percentile) platform global deviation, apex global deviation and angular deviation were 1.23 (0.9–1.4) mm/1.9 (1.2–2.3) mm ( p = .03, the Mann–Whitney U ‐test), 1.40 (1.1–1.6) mm/2.1 (1.7–3.9) mm ( p < .01) and 3.0 (0.9–6.0)°/6.7 (2.2–13.9)° ( p = .08), respectively. Both methods showed limited damage to the alveolar ridge and had similar peri‐ and post‐operative morbidity and safety.
Conclusions
Robot‐assisted implant placement enabled greater positional accuracy of the implant compared to freehand placement in this pilot trial. The robotic system should be further developed to simplify surgical procedures and improve accuracy and be validated in properly sized trials assessing the full spectrum of relevant outcomes.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Shanghai PerioImplant Innovation Center, Department of Oral and Maxillofacial Implantology, Shanghai Ninth People's Hospital, Shanghai Jiao Tong University School of Medicine, Shanghai, China.
- College of Stomatology, Shanghai Jiao Tong University, Shanghai, China.
- National Center for Stomatology, Shanghai, China.
- National Clinical Research Center for Oral Diseases, Shanghai, China.
- Shanghai Key Laboratory of Stomatology, Shanghai, China.
- European Research Group on Periodontology, Genoa, Italy.
ORCID-Profile: Xin-Yu Wu, Maurizio S Tonetti
Förderung
- Clinical Research Program of Ninth People's Hospital affiliated Shanghai Jiao Tong University School of Medicine
- National Clinical Research Center for Oral Diseases
- Shanghai Innovative Research Team Award of High-Level University
- Theta robot manufacturer (Jianjia Ltd., Hanghzhou, China)
Zitieren
APA
Shi, J. Y., Liu, B. L., Wu, X. Y., Liu, M., Zhang, Q., Lai, H. C., & Tonetti, M. S. (2024). Improved positional accuracy of dental implant placement using a haptic and machine‐vision‐controlled collaborative surgery robot: A pilot randomized controlled trial. Journal of Clinical Periodontology, 51(1), 24-32. https://doi.org/10.1111/jcpe.13893
Vancouver
Shi JY, Liu BL, Wu XY, Liu M, Zhang Q, Lai HC, et al. Improved positional accuracy of dental implant placement using a haptic and machine‐vision‐controlled collaborative surgery robot: A pilot randomized controlled trial. Journal of Clinical Periodontology. 2024;51(1):24-32. doi:10.1111/jcpe.13893
BibTeX
@article{Shi2024,
author = {Shi, Jun-Yu and Liu, Bei-Lei and Wu, Xin-Yu and Liu, Min and Zhang, Qi and Lai, Hong-Chang and Tonetti, Maurizio S},
title = {Improved positional accuracy of dental implant placement using a haptic and machine‐vision‐controlled collaborative surgery robot: A pilot randomized controlled trial},
journal = {Journal of Clinical Periodontology},
year = {2024},
volume = {51},
number = {1},
pages = {24--32},
doi = {10.1111/jcpe.13893}
}