Implantation Technique, Continuous Implant Geometry, and Longitudinal Low‐Voltage Impedance After Subcutaneous Implantable Cardioverter‐Defibrillator Implantation

Fabian Sanchis‐Gomar, Albert J. Rogers, Leah Schloss, Alexander C. Perino, Mohan N. Viswanathan, Tina Baykaner, Roy M. John, Nitish Badhwar, Sanjiv M. Narayan, Paul J. Wang, Marco V. Perez

Journal of Cardiovascular Electrophysiology · 17.11.2026

DOI: 10.1111/jce.70568 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Cardiovascular Electrophysiology, online erschienen am 17.11.2026. Verfasst von Fabian Sanchis‐Gomar und 10 weiteren Autor:innen. Die Arbeit nennt 18 Literaturangaben.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Defibrillation efficacy of the subcutaneous implantable cardioverter‐defibrillator (S‐ICD) depends on shock vector geometry and tissue impedances influenced by implantation technique. Whether longitudinal low‐voltage impedance (LVI) measurements differ across contemporary implantation strategies, and whether these differences are explained by quantitative radiographic geometry, remains incompletely characterized.

Objective

To determine whether S‐ICD implantation technique is associated with longitudinal LVI trajectories and whether continuous radiographic implant geometry provides information beyond categorical PRAETORIAN risk classification.

Methods

We performed a single‐center retrospective cohort study of 136 adults receiving an S‐ICD between 2014 and 2024, classified by operative documentation as subcutaneous (SC; n = 27), fluoroscopy‐marked intermuscular (FIM; n = 73), or fluoroscopy‐marked ultrasound‐guided intermuscular (FUSIM; n = 36) techniques. Per‐device LVI was summarized overall and across prespecified windows (0–30 days, 6 months–1 year, 2–5 years). Technique differences were analyzed. We fit a prespecified mixed‐effects interaction model that included coil–sternum distance, scaled in 5 mm increments, and assessed correlations between LVI and radiographic measures.

Results

In paired analyses, LVI increased by a mean of 13.3 Ω between the 0–30‐day and 2–5‐year windows. Overall LVI differed by implantation technique (Kruskal–Wallis p = 0.019), driven by higher values in SC versus FIM (Holm‐adjusted p = 0.041) and FUSIM ( p = 0.022). Additional impedance differences across techniques were absent in the 0–30‐day window but emerged during follow‐up. In the prespecified interaction mixed‐effects model, SC exhibited a significant additional rise relative to FUSIM during the 2–5‐year window (interaction β = 7.82 Ω, 95% CI 0.98−14.67; p = 0.025). Greater coil–sternum distance was independently associated with higher LVI (13.6 Ω per cm, 95% CI 4.2–23.0; p = 0.005) and correlated with chronic LVI ( n = 76; r = 0.441; Benjamini–Hochberg q < 0.001). The PRAETORIAN score was computable in 76 devices and classified 84% as low risk; within that low‐risk stratum, coil–sternum distance remained associated with chronic LVI ( ρ = 0.320, p = 0.010) whereas the score did not ( ρ = 0.136, p = 0.282), and each 5 mm of coil–sternum distance corresponded to 10.7 Ω higher LVI after adjustment for BMI (95% CI 2.68−18.73, p = 0.010). High‐voltage shock impedance at implant correlated strongly with chronic LVI ( n = 99; r = 0.71, p < 0.001), and the association persisted over 2–5 years ( n = 69; r = 0.61, p < 0.001).

Conclusion

Longitudinal LVI varies with S‐ICD implantation strategy, with time‐dependent divergence between SC and intermuscular implants. Within the low‐risk PRAETORIAN category, where the categorical score has no dynamic range, continuously measured coil–sternum distance remains associated with chronic impedance. Continuous implant geometry and longitudinal impedance complement, rather than replace, radiographic scoring of implant position.

Zitieren

APA

Sanchis‐Gomar, F., Rogers, A. J., Schloss, L., Perino, A. C., Viswanathan, M. N., Baykaner, T., John, R. M., Badhwar, N., Narayan, S. M., Wang, P. J., & Perez, M. V. (2026). Implantation Technique, Continuous Implant Geometry, and Longitudinal Low‐Voltage Impedance After Subcutaneous Implantable Cardioverter‐Defibrillator Implantation. Journal of Cardiovascular Electrophysiology. https://doi.org/10.1111/jce.70568

Vancouver

Sanchis‐Gomar F, Rogers AJ, Schloss L, Perino AC, Viswanathan MN, Baykaner T, et al. Implantation Technique, Continuous Implant Geometry, and Longitudinal Low‐Voltage Impedance After Subcutaneous Implantable Cardioverter‐Defibrillator Implantation. Journal of Cardiovascular Electrophysiology. 2026. doi:10.1111/jce.70568

BibTeX
@article{SanchisGomar2026,
  author = {Sanchis‐Gomar, Fabian and Rogers, Albert J. and Schloss, Leah and Perino, Alexander C. and Viswanathan, Mohan N. and Baykaner, Tina and John, Roy M. and Badhwar, Nitish and Narayan, Sanjiv M. and Wang, Paul J. and Perez, Marco V.},
  title = {Implantation Technique, Continuous Implant Geometry, and Longitudinal Low‐Voltage Impedance After Subcutaneous Implantable Cardioverter‐Defibrillator Implantation},
  journal = {Journal of Cardiovascular Electrophysiology},
  year = {2026},
  doi = {10.1111/jce.70568}
}
Kostenlos starten →