EUVIMEDEUVIMED entdecken
VergleichsstudieOpen Access

Advantages of the QDOT MICRO Catheter in Finding Conduction Gaps in Pulmonary Vein Isolation Compared With the Conventional 3.5‐mm Tip Electrode

Ken Masuda, Daigo Nagahara iD, Ryosuke Hatano iD, Kimitaka Nishizaki iD, Satoshi Yuda

Journal of Cardiovascular Electrophysiology · 10.08.2026

DOI: 10.1111/jce.70472 ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift Journal of Cardiovascular Electrophysiology, online erschienen am 10.08.2026. Verfasst von Ken Masuda und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiology, Teine Keijinkai Hospital, Sapporo, Hokkaido, Japan. Hauptthemen laut MEDLINE: Pulmonary Veins, Atrial Fibrillation, Catheter Ablation, Electrophysiologic Techniques, Cardiac. Die Arbeit nennt 17 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Pulmonary vein isolation (PVI) remains a cornerstone treatment for atrial fibrillation; however, residual conduction gaps often hinder first‐pass success and contribute to late reconnection, leading to arrhythmia recurrence. The QDOT MICRO catheter (Biosense Webster Inc., CA, USA) is equipped with three microelectrodes at its tip. Their small size and short inter‐electrode spacing allow for a more detailed recording of intracardiac potentials. This study evaluated the utility of the QDOT MICRO catheter in detecting conduction gaps during PVI.

Methods

Forty‐four gaps in 40 patients undergoing initial PVI were analyzed by comparing electrograms recorded using microelectrodes and conventional 3.5‐mm tip electrodes.

Results

At the gap sites, the local electrogram voltage recorded by microelectrodes was significantly higher than that by 3.5‐mm tip electrodes (median, 0.449 vs. 0.195 mV, p < 0.001), representing a 2.3‐fold amplification. The prevalence of fragmented potentials at the gap sites was significantly higher with microelectrodes than with 3.5‐mm tip electrodes (61% vs. 30%, p < 0.001). In 36% of gaps, fragmented potentials were detected with microelectrodes, whereas the 3.5‐mm electrodes recorded nonfragmented signals, contributing to accurate gap detection. In 11% of gaps, the amplitude was extremely low with 3.5‐mm tip electrodes [median 0.107 (IQR 0.088–0.112) mV], rendering the interpretation of local electrogram challenging. However, microelectrodes amplified the signal [median 0.215 (IQR 0.176–0.283) mV], allowing for accurate identification and interpretation.

Conclusion

Microelectrodes enhance the detection and characterization of conduction gaps, particularly under low‐voltage or in scarred tissue, thereby improving ablation accuracy.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Pulmonary VeinsAtrial FibrillationCatheter AblationElectrophysiologic Techniques, CardiacCardiac CathetersHeart Conduction SystemHumansMicroelectrodesEquipment DesignAction PotentialsFemaleMaleMiddle AgedPredictive Value of TestsAgedTreatment OutcomeMiniaturization

Stichwörter der Autor:innen

QDOT microconduction gaplocal electrogrammicro electrodepulmonary vein isolation

Autor:innen und Institutionen

  • Department of Cardiology, Teine Keijinkai Hospital, Sapporo, Hokkaido, Japan.

ORCID-Profile: Daigo Nagahara, Ryosuke Hatano, Kimitaka Nishizaki

Zitieren

APA

Masuda, K., Nagahara, D., Hatano, R., Nishizaki, K., & Yuda, S. (2026). Advantages of the QDOT MICRO Catheter in Finding Conduction Gaps in Pulmonary Vein Isolation Compared With the Conventional 3.5‐mm Tip Electrode. Journal of Cardiovascular Electrophysiology. https://doi.org/10.1111/jce.70472

Vancouver

Masuda K, Nagahara D, Hatano R, Nishizaki K, Yuda S. Advantages of the QDOT MICRO Catheter in Finding Conduction Gaps in Pulmonary Vein Isolation Compared With the Conventional 3.5‐mm Tip Electrode. Journal of Cardiovascular Electrophysiology. 2026. doi:10.1111/jce.70472

BibTeX
@article{Masuda2026,
  author = {Masuda, Ken and Nagahara, Daigo and Hatano, Ryosuke and Nishizaki, Kimitaka and Yuda, Satoshi},
  title = {Advantages of the QDOT MICRO Catheter in Finding Conduction Gaps in Pulmonary Vein Isolation Compared With the Conventional 3.5‐mm Tip Electrode},
  journal = {Journal of Cardiovascular Electrophysiology},
  year = {2026},
  doi = {10.1111/jce.70472}
}
Kostenlos starten →