Vergleichsstudie in der Zeitschrift Journal of Cardiovascular Electrophysiology, online erschienen am 02.03.2026. Verfasst von Auroa Badin und 9 weiteren Autor:innen. Beteiligte Einrichtung: Division of Cardiac Electrophysiology, OhioHealth Heart and Vascular Physicians, Riverside Methodist Hospital, Columbus, Ohio, USA. Hauptthemen laut MEDLINE: Atrial Fibrillation, Electrophysiologic Techniques, Cardiac, Imaging, Three-Dimensional, Pulmonary Veins. Die Arbeit nennt 5 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Introduction
Pulsed field ablation (PFA) is an established therapy for atrial fibrillation (AF), yet optimal mapping strategies during pentaspline PFA remain uncertain. We compared procedural characteristics and material costs of an integrated 3D electroanatomic mapping (EAM) platform with conventional mapping systems. Methods and Results In a multicenter observational study of 448 consecutive AF ablations done using pentaspline catheter (62 integrated mapping; 386 conventional mapping) performed between February 2024 and August 2025, baseline characteristics were similar between groups. Compared with conventional mapping, the integrated platform was associated with shorter total procedure time (49 ± 16 vs. 66 ± 26 min, p < 0.001), shorter left atrial dwell time (35 ± 12 vs. 46 ± 30 min, p < 0.001), reduced fluoroscopy time (6 ± 4 vs. 12 ± 7 min, p < 0.001), lower radiation dose (36 ± 10 vs. 93 ± 18 mGy, p < 0.001), and lower mean material cost ($12 958 ± 164 vs. $14 739 ± 1192, p < 0.001). Differences were consistent across lesion sets. Acute pulmonary vein isolation was achieved in all patients, with two major complications occurring in the conventional mapping group.
Conclusion
Integrated 3D EAM during pentaspline PFA was associated with shorter procedures, lower radiation exposure, and reduced material costs. These findings support the procedural and operational advantages of integrated mapping platforms in contemporary AF ablation practice.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Division of Cardiac Electrophysiology, OhioHealth Heart and Vascular Physicians, Riverside Methodist Hospital, Columbus, Ohio, USA.
- Department of Immunity and Inflammation, Cleveland Clinic Research, Cleveland, Ohio, USA.
ORCID-Profile: Auroa Badin
Zitieren
APA
Badin, A., Robinson, A. K., Farrah, A., Billakanty, S., Nemer, D. M., Shah, A. N., Tyler, J. D., Fu, E. Y., Chopra, N., & Amin, A. K. (2026). Comparison of Integrated 3D Electroanatomic Mapping and Conventional Mapping Systems for Pentaspline Pulsed Field Ablation of Atrial Fibrillation. Journal of Cardiovascular Electrophysiology, 37(6), 1353-1355. https://doi.org/10.1111/jce.70292
Vancouver
Badin A, Robinson AK, Farrah A, Billakanty S, Nemer DM, Shah AN, et al. Comparison of Integrated 3D Electroanatomic Mapping and Conventional Mapping Systems for Pentaspline Pulsed Field Ablation of Atrial Fibrillation. Journal of Cardiovascular Electrophysiology. 2026;37(6):1353-1355. doi:10.1111/jce.70292
BibTeX
@article{Badin2026,
author = {Badin, Auroa and Robinson, Andrea K and Farrah, Allyson and Billakanty, Saanvi and Nemer, David M and Shah, Ankur N and Tyler, Jaret D and Fu, Eugene Y and Chopra, Nagesh and Amin, Anish K},
title = {Comparison of Integrated 3D Electroanatomic Mapping and Conventional Mapping Systems for Pentaspline Pulsed Field Ablation of Atrial Fibrillation},
journal = {Journal of Cardiovascular Electrophysiology},
year = {2026},
volume = {37},
number = {6},
pages = {1353--1355},
doi = {10.1111/jce.70292}
}
EUVIMED