Role of a Deep‐Learning Based Convolutional Neural Network Model for Real‐Time Ventricular Tachycardia Alarm Classification
Journal of Cardiovascular Electrophysiology · 2026
DOI: 10.1111/jce.70271 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Cardiovascular Electrophysiology, 2026. Verfasst von Unmesh Khanolkar und 5 weiteren Autor:innen. Beteiligte Einrichtung: Division of Cardiac Electrophysiology, University of Toledo, Toledo, Ohio, USA. Hauptthemen laut MEDLINE: Convolutional Neural Networks, Tachycardia, Ventricular, Deep Learning, Signal Processing, Computer-Assisted. Die Arbeit nennt 22 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background
Ventricular tachycardia (VT) is a life‐threatening arrhythmia that requires both relatively rapid and accurate detection in intensive care units (ICUs). Continuous monitoring systems play a crucial role in detecting them. However, previous studies have reported that nearly 9 out of 10 arrhythmia alarms in ICUs tend to be false positives, which usually transpire to a well‐documented phenomenon called “alarm fatigue” that leads to desensitization, delayed responses, and increased cognitive burden on healthcare providers.
Methods
We developed a deep learning based, one‐dimensional convolutional neural network (1D‐CNN) to classify VT alarms using multiple raw waveform inputs, including two electrocardiogram (ECG) leads, photoplethysmogram (PPG) and arterial blood pressure (ABP) signals. The model was trained using the publicly available VTaC Arrhythmia Benchmark Dataset. We used the 10‐second waveform segments that preceded each VT alarm, pre‐processed and then used them to train the machine learning model to correctly classify the VT alarm.
Results
On the test set, the model achieved an area under the receiver operating characteristic curve of 0.901, overall accuracy of 83.22%, F1‐score of 73.3%, sensitivity of 77.53%, specificity of 85.63%, and positive predictive value of 69.57%. The model successfully detected over three‐quarters of them while significantly reducing false positive rates for the detection of VT.
Conclusions
This study demonstrates that a deep learning based 1D‐CNN model using short segments of raw waveform data can achieve robust performance in distinguishing true and false VT alarms.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Division of Cardiac Electrophysiology, University of Toledo, Toledo, Ohio, USA.
- NYU Tandon School of Engineering, New York, New York, USA.
- Indian Institute of Technology, Mandi, India.
- Department of Bioengineering, University of Toledo, Toledo, Ohio, USA.
- Department of Anaesthesia, University of Toledo, Toledo, Ohio, USA.
- Mayo Clinic, Rochester, Minnesota, USA.
Zitieren
APA
Khanolkar, U., Yadav, A., Mann, A., Pappada, S., Deshmukh, A., & Maan, A. (2026). Role of a Deep‐Learning Based Convolutional Neural Network Model for Real‐Time Ventricular Tachycardia Alarm Classification. Journal of Cardiovascular Electrophysiology. https://doi.org/10.1111/jce.70271
Vancouver
Khanolkar U, Yadav A, Mann A, Pappada S, Deshmukh A, Maan A. Role of a Deep‐Learning Based Convolutional Neural Network Model for Real‐Time Ventricular Tachycardia Alarm Classification. Journal of Cardiovascular Electrophysiology. 2026. doi:10.1111/jce.70271
BibTeX
@article{Khanolkar2026,
author = {Khanolkar, Unmesh and Yadav, Ashish and Mann, Avdhesh and Pappada, Scott and Deshmukh, Abhishek and Maan, Abhishek},
title = {Role of a Deep‐Learning Based Convolutional Neural Network Model for Real‐Time Ventricular Tachycardia Alarm Classification},
journal = {Journal of Cardiovascular Electrophysiology},
year = {2026},
doi = {10.1111/jce.70271}
}
EUVIMED