Beobachtungsstudie in der Zeitschrift Journal of Cardiovascular Electrophysiology, 2026. Verfasst von Alexander Smith und 3 weiteren Autor:innen. Beteiligte Einrichtung: Cooper Medical School of Rowan University, Camden, New Jersey, USA. Hauptthemen laut MEDLINE: Guideline Adherence, Defibrillators, Implantable, Practice Guidelines as Topic, Pacemaker, Artificial. Die Arbeit nennt 16 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Introduction

Despite data‐driven 2015 expert consensus statement (consensus) guidelines on recommended PPM and ICD parameters, patients often fall short of optimal device programing. The purpose of this retrospective, observational, single‐center study is to review device implant registry data both before and after the consensus update to determine whether there is an improvement in adherence to consensus‐recommended device programming over time.

Methods

McKesson software was used to identify all patients pre‐ (1/1/2012–12/31/2013) and post‐ (1/1/2020–12/31/2022) consensus who underwent a single, dual, or BiV device implant and had at least one follow‐up appointment in the electronic medical record. Groups were evaluated for adherence to consensus recommendations based on the type of device. BiV devices were evaluated based on the percent of total pacing, with > 98% pacing being considered compliant with the consensus. Single and dual‐chamber ICDs were evaluated based on bradycardia and tachyarrhythmia programming parameters as outlined in the 2015 consensus document and the 2019 manufacturer‐specific update. Due to significant implantation rates at our institution, and to maintain consistency, only Medtronic devices were evaluated in this study.

Results

In the BiV group, before the consensus, 41 (85.4%) patients were > 98% paced, compared with only 69 (58.0%) patients post‐consensus. Forty‐five (93.7%) patients were at least 90% paced in the pre‐consensus group, compared with 109 (91.6%) in the post‐consensus group. In the ICD group, bradycardia‐pacing settings were similar in both groups (34/89.5% vs. 73/90.1%, p = 0.91). There was a trend toward improved adherence to tachyarrhythmia therapies post‐consensus (15/39.5% vs. 37/45.7%, p = 0.52). Nonadherence was driven by 1:1 SVT logic initiation. If this parameter was excluded, both groups increased significantly (37/97.4% vs. 78/96.3%, p = 0.76).

Conclusions

This study adds to the growing body of literature showing a persistent disconnect between data‐driven recommendations and real‐world clinical practice. Larger studies are needed to better understand physician barriers to these guideline‐driven recommendations.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Guideline AdherenceDefibrillators, ImplantablePractice Guidelines as TopicPacemaker, ArtificialEvidence-Based MedicineElectric CountershockCardiac Resynchronization Therapy DevicesCardiac Resynchronization TherapyCardiac Pacing, ArtificialArrhythmias, CardiacPractice Patterns, Physicians'HumansRetrospective StudiesRegistriesTreatment OutcomeTime FactorsFemaleMaleAgedMiddle AgedConsensus

Stichwörter der Autor:innen

BiV pacingICD programmingadherence

Autor:innen und Institutionen

  • Cooper Medical School of Rowan University, Camden, New Jersey, USA.

ORCID-Profile: Alexander Smith, Matthew Ortman

Zitieren

APA

Smith, A., Ortman, M., Andriulli, J., & Russo, A. M. (2026). Pacemaker and ICD Programming Adherence With Evidence‐Based Recommendations. Journal of Cardiovascular Electrophysiology. https://doi.org/10.1111/jce.70242

Vancouver

Smith A, Ortman M, Andriulli J, Russo AM. Pacemaker and ICD Programming Adherence With Evidence‐Based Recommendations. Journal of Cardiovascular Electrophysiology. 2026. doi:10.1111/jce.70242

BibTeX
@article{Smith2026,
  author = {Smith, Alexander and Ortman, Matthew and Andriulli, John and Russo, Andrea M},
  title = {Pacemaker and ICD Programming Adherence With Evidence‐Based Recommendations},
  journal = {Journal of Cardiovascular Electrophysiology},
  year = {2026},
  doi = {10.1111/jce.70242}
}
Kostenlos starten →