Elective Generator Replacement of the Subcutaneous Implantable Defibrillator—Always a Simple Pit Stop?

Fabian Fastenrath, Sara Wuerfel, Volker Liebe, Ibrahim Akin, Daniel Duerschmied, Martin Borggrefe, Juergen Kuschyk iD, Boris Rudic

Journal of Cardiovascular Electrophysiology · 2025

DOI: 10.1111/jce.16638 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Cardiovascular Electrophysiology, 2025. Verfasst von Fabian Fastenrath und 7 weiteren Autor:innen. Beteiligte Einrichtung: Cardiology, Angiology, Haemostaseology, and Medical Intensive Care, Medical Centre Mannheim, Medical Faculty Mannheim, Heidelberg University, Mannhei…. Hauptthemen laut MEDLINE: Defibrillators, Implantable, Electric Countershock, Device Removal, Arrhythmias, Cardiac. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 19 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: The authors received no specific funding for this work..

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background and Aims The safety and efficacy of the subcutaneous implantable cardioverter‐defibrillator (S‐ICD) has been proven in various clinical trials. Data on device replacement strategies is scarce. This study aims to evaluate long‐term shock efficacy, trends in shock impedance, and the effects of a replacement strategy at elective device replacement.

Methods

Between January 2018 and November 2022, a total of 99 consecutive patients underwent S‐ICD device replacement. Shock impedance was determined at initial implantation, before and after device replacement. In case of high shock impedance, additional optimization of lead and generator position was performed if applicable, as well as the removal of fibrous encapsulation (“capsulectomy”) of the S‐ICD pocket.

Results

Defibrillation testing with the first 65 J shock was successful in 87% of patients after initial device implantation versus 85% with the same device immediately before device replacement ( p = 0.88). Eight patients preoperatively failed defibrillation testing with 65 J and 80 J. After a mean dwell time of 65 ± 21 months, shock impedance increased significantly from 82 ± 23 Ω at initial implantation to 98 ± 37 Ω at replacement ( p = 0.004). Capsulectomy and optimized repositioning of the S‐ICD lead and generator resulted in a significant decrease of shock impedance to 81 ± 25 Ω as compared to 98 ± 37 Ω before the replacement ( p = 0.004). 65 J shock efficacy improved to 95%. Postoperative defibrillation testing was successful in all patients. No acute complications were observed.

Conclusion

High conversion rates at device replacement are achievable through optimized replacement strategy. In addition to improving the lead and generator position, capsulectomy appears to be a safe and effective way to reduce shock impedance and might contribute to overall system performance.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Defibrillators, ImplantableElectric CountershockDevice RemovalArrhythmias, CardiacProsthesis ImplantationHumansMaleFemaleMiddle AgedAgedTreatment OutcomeTime FactorsRetrospective StudiesElectric ImpedanceRisk FactorsProsthesis Design

Stichwörter der Autor:innen

S‐ICDdefibrillation efficacydefibrillation failuregenerator replacementimplantable cardioverter‐defibrillatorshock impedancesubcutaneous defibrillator

Autor:innen und Institutionen

  • Cardiology, Angiology, Haemostaseology, and Medical Intensive Care, Medical Centre Mannheim, Medical Faculty Mannheim, Heidelberg University, Mannheim, Germany.
  • European Center for AngioScience (ECAS), German Center for Cardiovascular Research (DZHK) partner site Heidelberg/Mannheim, and Centre for Cardiovascular Acute Medicine Mannheim (ZKAM), Medical Centre Mannheim, Heidelbe…

ORCID-Profile: Juergen Kuschyk

Förderung

  • The authors received no specific funding for this work.

Zitieren

APA

Fastenrath, F., Wuerfel, S., Liebe, V., Akin, I., Duerschmied, D., Borggrefe, M., Kuschyk, J., & Rudic, B. (2025). Elective Generator Replacement of the Subcutaneous Implantable Defibrillator—Always a Simple Pit Stop?. Journal of Cardiovascular Electrophysiology. https://doi.org/10.1111/jce.16638

Vancouver

Fastenrath F, Wuerfel S, Liebe V, Akin I, Duerschmied D, Borggrefe M, et al. Elective Generator Replacement of the Subcutaneous Implantable Defibrillator—Always a Simple Pit Stop?. Journal of Cardiovascular Electrophysiology. 2025. doi:10.1111/jce.16638

BibTeX
@article{Fastenrath2025,
  author = {Fastenrath, Fabian and Wuerfel, Sara and Liebe, Volker and Akin, Ibrahim and Duerschmied, Daniel and Borggrefe, Martin and Kuschyk, Juergen and Rudic, Boris},
  title = {Elective Generator Replacement of the Subcutaneous Implantable Defibrillator—Always a Simple Pit Stop?},
  journal = {Journal of Cardiovascular Electrophysiology},
  year = {2025},
  doi = {10.1111/jce.16638}
}
Kostenlos starten →