Transvenous extraction of conduction system and lumenless pacing leads
Journal of Cardiovascular Electrophysiology · 2024
DOI: 10.1111/jce.16467 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift Journal of Cardiovascular Electrophysiology, 2024. Verfasst von Ethan S Wagner und 5 weiteren Autor:innen. Beteiligte Einrichtung: Duke University School of Medicine, Durham, North Carolina, USA. Hauptthemen laut MEDLINE: Cardiac Pacing, Artificial, Pacemaker, Artificial, Device Removal. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 31 Literaturangaben.
Abstract (Originalsprache)
Introduction
Conduction system pacing (CSP), often accomplished with lumenless pacing leads, is increasingly employed to achieve physiologic ventricular activation. There are limited data on the extraction of these leads. The objective of this study was to describe the safety and efficacy of extraction of CSP pacing leads and compare outcomes with extraction of non‐CSP lumenless leads.
Methods
Patients undergoing CSP/non‐CSP lumenless lead removal were included. Outcomes of interest included rates of complete procedural success, complications, and successful reimplantation.
Results
Overall, 23 patients were included ( n = 14 with CSP and n = 9 with non‐CSP lumenless leads implanted in the right atrium, right ventricle, or septum). The mean age was 52.7 ± 24.0 years, 30% were female, and the mean lead age was 4.5 ± 4.4 years. The complete procedural success rate was 100%. One serious complication occurred in the non‐CSP group but was unrelated to the lead of interest. Manual traction alone was successful in 57% of CSP cases (mean lead age 2.4 ± 1.7 years) and in 11% of non‐CSP cases (mean lead age 7.9 ± 5.3 years). Laser sheaths were used in 43% of CSP cases and 89% of non‐CSP cases; rotational cutting tools were used in no CSP cases and in 33% of non‐CSP cases. Reimplantation in the conduction system was attempted with a left bundle branch pacing lead and successful in 80% ( n = 4/5).
Conclusion
Extraction of CSP and non‐CSP lumenless leads is feasible with a high success rate and a good safety profile. CSP reimplantation after extraction is also feasible with good electrical performance.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Duke University School of Medicine, Durham, North Carolina, USA.
- Clinical Cardiac Electrophysiology Section, Division of Cardiology, Department of Medicine, Duke University Hospital, Durham, North Carolina, USA.
ORCID-Profile: Ethan S Wagner, Robert K Lewis, Daniel J Friedman, Jonathan P Piccini
Zitieren
APA
Wagner, E. S., Lewis, R. K., Pokorney, S. D., Hegland, D. D., Friedman, D. J., & Piccini, J. P. (2024). Transvenous extraction of conduction system and lumenless pacing leads. Journal of Cardiovascular Electrophysiology. https://doi.org/10.1111/jce.16467
Vancouver
Wagner ES, Lewis RK, Pokorney SD, Hegland DD, Friedman DJ, Piccini JP. Transvenous extraction of conduction system and lumenless pacing leads. Journal of Cardiovascular Electrophysiology. 2024. doi:10.1111/jce.16467
BibTeX
@article{Wagner2024,
author = {Wagner, Ethan S and Lewis, Robert K and Pokorney, Sean D and Hegland, Donald D and Friedman, Daniel J and Piccini, Jonathan P},
title = {Transvenous extraction of conduction system and lumenless pacing leads},
journal = {Journal of Cardiovascular Electrophysiology},
year = {2024},
doi = {10.1111/jce.16467}
}
EUVIMED