Voltage and propagation mapping: New tools to improve successful ablation of atrioventricular nodal reentry tachycardia
Journal of Cardiovascular Electrophysiology · 2024
DOI: 10.1111/jce.16234 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Cardiovascular Electrophysiology, 2024. Verfasst von Chiara Devecchi und 9 weiteren Autor:innen. Beteiligte Einrichtung: Division of Cardiology, Ospedale Sant'Andrea, Vercelli, Italy. Hauptthemen laut MEDLINE: Tachycardia, Atrioventricular Nodal Reentry, Catheter Ablation, Action Potentials, Electrophysiologic Techniques, Cardiac. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 20 Literaturangaben.
Abstract (Originalsprache)
Introduction
Mapping system is useful in ablation of atrioventricular nodal reentry tachycardia (AVNRT) and localization of anatomic variances. Voltage mapping identifies a low voltage area in the Koch triangle called low‐voltage‐bridge (LVB); propagation mapping identifies the collision point (CP) of atrial wavefront convergence. We conducted a prospective study to evaluate the relationship between LVB and CP with successful site of ablation and identify standard value for LVB. Materials and Methods Three‐dimensional (3D) maps of the right atria were constructed from intracardiac recordings using the ablation catheter. Cut‐off values on voltage map were adjusted until LVB was observed. On propagation map, atrial wavefronts during sinus rhythm collide in the site representing CP, indicating the area of slow pathway conduction. Ablation site was selected targeting LVB and CP site, confirmed by anatomic position on fluoroscopy and atrioventricular ratio.
Results
Twenty‐seven consecutive patients were included. LVB and CP were present in all patients. Postprocedural evaluation identified standard cut‐off of 0.3−1 mV useful for LVB identification. An overlap between LVB and CP was observed in 23 (85%) patients. Procedure success was achieved in all patient with effective site at first application in 22 (81%) patients. There was a significant correlation between LVB, CP, and the site of effective ablation ( p = .001).
Conclusion
We found correlation between LVB and CP with the site of effective ablation, identifying a voltage range useful for standardized LVB identification. These techniques could be useful to identify ablation site and minimize radiation exposure.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Division of Cardiology, Ospedale Sant'Andrea, Vercelli, Italy.
- Division of Cardiology, AOU Maggiore della Carità, Novara, Italy.
- Division of Cardiology, AOU Città della Salute e della Scienza, Torino, Italy.
- Abbott S.R.L., Roma, Italy.
- Department of Translational Medicine, University of Eastern Piedmont, Novara, Italy.
ORCID-Profile: Chiara Devecchi, Massimo Magnano
Zitieren
APA
Devecchi, C., Matta, M., Magnano, M., Dell'Era, G., Galiffa, V. A., Renaudo, D., Negro, A., Occhetta, E., Patti, G., & Rametta, F. (2024). Voltage and propagation mapping: New tools to improve successful ablation of atrioventricular nodal reentry tachycardia. Journal of Cardiovascular Electrophysiology. https://doi.org/10.1111/jce.16234
Vancouver
Devecchi C, Matta M, Magnano M, Dell'Era G, Galiffa VA, Renaudo D, et al. Voltage and propagation mapping: New tools to improve successful ablation of atrioventricular nodal reentry tachycardia. Journal of Cardiovascular Electrophysiology. 2024. doi:10.1111/jce.16234
BibTeX
@article{Devecchi2024,
author = {Devecchi, Chiara and Matta, Mario and Magnano, Massimo and Dell'Era, Gabriele and Galiffa, Vincenzo Alessandro and Renaudo, Dario and Negro, Andrea and Occhetta, Eraldo and Patti, Giuseppe and Rametta, Francesco},
title = {Voltage and propagation mapping: New tools to improve successful ablation of atrioventricular nodal reentry tachycardia},
journal = {Journal of Cardiovascular Electrophysiology},
year = {2024},
doi = {10.1111/jce.16234}
}
EUVIMED